[Paper Review] Comparison of an open-hardware electroencephalography amplifier with medical grade device in brain-computer interface applications

Jérémy Frey|arXiv (Cornell University)|Jun 8, 2016
EEG and Brain-Computer InterfacesNeuroscience8 references6 citations
TL;DR

This study compares the open-source OpenBCI EEG amplifier with the medical-grade g.tec g.USBamp in parallel recordings using identical electrodes and setups. Despite lower sampling rates and lack of medical certification, OpenBCI achieved classification accuracy close to the g.USBamp in both P300 speller and workload monitoring tasks, suggesting it is a viable, low-cost alternative for non-medical BCI applications.

ABSTRACT

Brain-computer interfaces (BCI) are promising communication devices between humans and machines. BCI based on non-invasive neuroimaging techniques such as electroencephalography (EEG) have many applications , however the dissemination of the technology is limited, in part because of the price of the hardware. In this paper we compare side by side two EEG amplifiers, the consumer grade OpenBCI and the medical grade g.tec g.USBamp. For this purpose, we employed an original montage, based on the simultaneous recording of the same set of electrodes. Two set of recordings were performed. During the first experiment a simple adapter with a direct connection between the amplifiers and the electrodes was used. Then, in a second experiment, we attempted to discard any possible interference that one amplifier could cause to the other by adding "ideal" diodes to the adapter. Both spectral and temporal features were tested -- the former with a workload monitoring task, the latter with an visual P300 speller task. Overall, the results suggest that the OpenBCI board -- or a similar solution based on the Texas Instrument ADS1299 chip -- could be an effective alternative to traditional EEG devices. Even though a medical grade equipment still outperforms the OpenBCI, the latter gives very close EEG readings, resulting in practice in a classification accuracy that may be suitable for popularizing BCI uses.

Motivation & Objective

  • To evaluate whether open-hardware EEG amplifiers like OpenBCI can serve as a cost-effective alternative to medical-grade EEG systems in BCI applications.
  • To address the limitation of prior studies that compared EEG systems under non-identical conditions by enabling simultaneous recording from the same electrodes.
  • To investigate the impact of electrical interference between two amplifiers when connected in parallel using a custom adapter.
  • To assess the performance of OpenBCI in two distinct BCI tasks: one emphasizing temporal features (P300 speller) and one focusing on spectral features (workload monitoring).

Proposed method

  • Designed a custom printed circuit board (PCB) adapter to connect both the OpenBCI and g.USBamp amplifiers simultaneously to the same set of 16 g.Ladybird active electrodes.
  • Conducted two experimental conditions: first with direct parallel connection of amplifiers, second with an adapter incorporating ideal diodes to isolate signal paths and prevent cross-interference.
  • Used a standardized montage with wet, active electrodes from g.tec to ensure consistent signal input across both amplifiers.
  • Recorded EEG data during two tasks: a visual P300 speller task to evaluate temporal event-related potential (ERP) features and a workload monitoring task to assess spectral power changes.
  • Computed classification accuracy using a linear classifier on both temporal (ERP) and spectral (power band) features from each amplifier.
  • Measured signal correlation between the two amplifiers using Pearson correlation (R) on averaged ERPs and spectral power, and evaluated performance via AUROCC scores.

Experimental results

Research questions

  • RQ1Can the OpenBCI EEG amplifier produce EEG recordings of sufficient quality to support reliable BCI performance when compared to a medical-grade device like the g.USBamp?
  • RQ2Does the presence of a second amplifier in parallel introduce measurable signal distortion or interference that affects recording quality?
  • RQ3How do the classification accuracies of OpenBCI and g.USBamp compare across different BCI tasks—specifically, one relying on temporal features (P300) and another on spectral features (workload)?
  • RQ4To what extent do the EEG signals recorded by OpenBCI correlate with those from the g.USBamp under identical physiological conditions?
  • RQ5Can the use of ideal diodes in the adapter effectively isolate the amplifiers and reduce interference, thereby improving measurement fidelity?

Key findings

  • The OpenBCI amplifier achieved AUROCC scores well above chance level in both the P300 speller and workload monitoring tasks, indicating reliable classification performance.
  • The correlation between averaged ERPs recorded by OpenBCI and g.USBamp reached a mean R score of 0.88 in the isolated diode setup, indicating strong signal similarity.
  • Despite lower sampling rate (125Hz vs. up to 16,000Hz possible on ADS1299), OpenBCI's performance was comparable to the g.USBamp in the tested BCI applications.
  • The direct connection setup showed slightly lower ERP correlation (R ≈ 0.88) than expected, suggesting possible interference from the adapter, though classification accuracy remained high.
  • The use of ideal diodes in the adapter reduced cross-talk and improved signal fidelity, though no significant performance difference was observed between the two adapter types.
  • OpenBCI’s EEG readings were very close to those of the g.USBamp, with classification accuracy in practice suitable for non-medical BCI applications, despite lacking medical certification.