[Paper Review] Compressive Raman microspectroscopy parallelized by single-photon avalanche diode arrays
This paper presents a compressive Raman microspectroscopy system that leverages a 512-pixel LinoSPAD2 single-photon avalanche diode array to parallelize detection, achieving an effective pixel dwell time of 23 µs—representing over 10× speedup over prior single-pixel approaches. By combining line-scanning illumination with programmable spectral filtering via a DMD and high-throughput SPAD array, the method enables real-time, high-speed chemical imaging with minimal noise and scalable performance, paving the way for video-rate spontaneous Raman imaging.
We demonstrate an efficient and scalable compressive Raman parallelization scheme based on single-photon avalanche diode (SPAD) arrays to reach pixel dwell times of 23 $μ$s, representing over 10$ imes$ speed-up using the otherwise weak spontaneous Raman effect.
Motivation & Objective
- Overcome the fundamental speed limitations of conventional Raman microspectroscopy due to weak spontaneous Raman scattering.
- Address the bottleneck of single-pixel detection in compressive Raman by introducing spatial parallelization using SPAD arrays.
- Enable real-time, high-speed hyperspectral imaging with minimal noise and high sensitivity using low-light, low-noise SPAD technology.
- Minimize spectral crosstalk and optimize signal fidelity in a programmable spectrometer by engineering the point spread function (PSF) and managing DMD dispersion.
- Demonstrate a scalable, cost-effective platform for video-rate, label-free chemical imaging in biological and materials science applications.
Proposed method
- Employ a programmable spectrometer based on a digital micromirror device (DMD) to route specific spectral bins to a single-pixel SPAD, enabling compressive sensing of hyperspectral data.
- Integrate a high-throughput, 512-pixel LinoSPAD2 linear array with 1:1 direct readout via custom FPGA boards to parallelize detection across spatial pixels.
- Use line-scanning illumination (90 µm wide) focused via a 20× objective (NA 0.5) to excite the sample, with 1D confocal detection through a 30 µm slit to limit background noise.
- Implement a 1D scanning stage to raster the excitation line across the sample, acquiring data in a line-by-line fashion with adjustable line dwell time (τldt).
- Optimize the spectrometer's point spread function (PSF) to minimize inter-pixel spectral crosstalk, achieving PSFs down to tens of µm for effective coupling with SPAD pixels.
- Apply Fourier-domain Gaussian smoothing for denoising and reconstruct 3D chemical images (x, y, z) from compressed spectral measurements using supervised compressive Raman algorithms.

Experimental results
Research questions
- RQ1Can spatial parallelization using a SPAD array significantly reduce pixel dwell time in compressive Raman microspectroscopy compared to single-pixel detection?
- RQ2To what extent can DMD-induced spectral dispersion be mitigated to enable high-fidelity, low-crosstalk detection with SPAD arrays?
- RQ3What is the achievable effective pixel dwell time (τpdt) using a parallelized compressive Raman setup with a high-throughput SPAD array?
- RQ4How does the detection efficiency (PDE) of the SPAD array compare to single-pixel SPADs, and what improvements are possible with optical enhancements?
- RQ5Can this system achieve video-rate imaging speeds in spontaneous Raman microspectroscopy using only low excitation power and standard laser sources?
Key findings
- An effective pixel dwell time of 23 µs was achieved using a 128-pixel subset of the 512-pixel LinoSPAD2 array, representing a 10× improvement over previous single-pixel compressive Raman systems.
- The system demonstrated a detection efficiency (PDE) of 7% for the SPAD array, with potential to increase to 20% using microlens integration, enabling a 3× improvement in sensitivity.
- Using all 512 SPAD pixels instead of 128 could provide an additional 4× gain in acquisition speed, while utilizing both DMD outputs could yield a further 2× speedup.
- The point spread function (PSF) was engineered to be stable and compact (tens of µm), minimizing inter-pixel spectral crosstalk and enabling high-fidelity spectral detection across the array.
- The system achieved 3D chemical imaging of polystyrene (6 µm) and PMMA (5 µm) beads in water at 23 µs effective dwell time, with image quality preserved after denoising via Fourier-domain Gaussian smoothing.
- Theoretical analysis suggests that further improvements could reduce the effective pixel dwell time by up to 20×, approaching the fundamental limit of the detector's dead time, enabling true video-rate Raman imaging.