[Paper Review] DECal: A Spectrophotometric Calibration System For DECam

John Marshall, Jean-Philippe Rheault|arXiv (Cornell University)|Feb 22, 2013
Advanced Chemical Sensor TechnologiesEngineering1 references5 citations
TL;DR

DECal is a spectrophotometric calibration system for the DECam instrument on the CTIO 4m Blanco telescope, using a monochromator-based tunable light source and a Lambertian flat field screen to measure the wavelength-dependent instrumental response across 300–1100 nm. It enables monthly monitoring of the telescope+instrument throughput, achieving 1% photometric accuracy by correcting for filter transmission changes, coating degradation, and system throughput variations.

ABSTRACT

DECal is a new calibration system for the CTIO 4 m Blanco telescope. It is currently being installed as part of the Dark Energy Survey and will provide both broadband flat fields and narrowband (about 1 nm bandwidth) spectrophotometric calibration for the new Dark Energy Camera (DECam). Both of these systems share a new Lambertian flat field screen. The broadband flat field system uses LEDs to illuminate each photometric filter. The spectrophotometric calibration system consists of a monochromator-based tunable light source that is projected onto the flat field screen using a custom line-to-spot fiber bundle and an engineered diffuser. Several calibrated photodiodes positioned along the beam monitor the telescope throughput as a function of wavelength. This system will measure the wavelength-dependent instrumental response function of the total telescope+instrument system in the range 300 < lambda < 1100nm. The spectrophotometric calibration will be performed regularly (roughly once per month) to determine the spectral response of the DECam system and to monitor changes in instrumental throughput during the five year Dark Energy Survey project.

Motivation & Objective

  • To enable highly accurate and precise photometry for the Dark Energy Survey (DES) by correcting instrumental and atmospheric effects.
  • To address the challenge of maintaining photometric accuracy over five years by monitoring changes in the telescope+instrument response function.
  • To provide a stable, low-maintenance calibration system suitable for long-term use in routine astronomical surveys.
  • To measure the true spectral response of DECam filters, especially those with incomplete or uncertain manufacturer data, such as the u-band filter.
  • To support improved supernova k-corrections and photometric redshifts through precise knowledge of filter transmission functions.

Proposed method

  • A monochromator-based tunable light source generates narrowband (≈1 nm) illumination across 300–1100 nm, replacing tunable lasers for lower maintenance.
  • Light is coupled into a custom line-to-spot fiber bundle and projected onto a Lambertian flat field screen to simulate uniform sky illumination.
  • Calibrated photodiodes along the beam path monitor the telescope's throughput as a function of wavelength, enabling measurement of the full instrumental response function.
  • The system uses a single, large, lightweight aluminum honeycomb flat field screen coated with a highly reflective, Lambertian material to ensure uniform illumination across the telescope's field of view.
  • Broadband flat fields are produced using LEDs to illuminate each photometric filter, enabling daily calibration of pixel-to-pixel response variations in DECam CCDs.
  • The system is designed for regular, automated operation—monthly spectrophotometric scans and daily flat fields—during the five-year DES survey.

Experimental results

Research questions

  • RQ1How can the wavelength-dependent instrumental response of the DECam+telescope system be measured with high precision and stability over five years?
  • RQ2What is the true spectral transmission function of DECam filters, particularly those with incomplete or uncertain manufacturer specifications?
  • RQ3Can a monochromator-based system provide sufficient signal-to-noise and stability for routine, long-term calibration in optical and infrared bands?
  • RQ4How do changes in optical coatings and filter performance affect instrumental throughput, and can they be monitored effectively with this system?
  • RQ5To what extent can this calibration system improve photometric accuracy for supernova k-corrections and photometric redshift measurements?

Key findings

  • The DECal prototype successfully measured the true spectral response of the u-band filter for the Carnegie Supernova Project, correcting an initial assumption that led to photometric inaccuracies.
  • Spectrophotometric scans of optical and infrared filters on the Swope and du Pont telescopes achieved 1% accuracy in throughput measurements, validating the system’s precision.
  • The system demonstrated feasibility for infrared calibration, with measurable throughput in the J, H, and Ks bands, though signal-to-noise was lower at the red end of the Ks band.
  • The monochromator-based light source proved effective and required less maintenance than tunable lasers, supporting long-term deployment.
  • The Lambertian flat field screen ensured uniform illumination across the telescope’s field of view, enabling accurate flat-field correction and throughput calibration.
  • The system is scalable and can be extended into the near-infrared (up to 2400 nm) with minimal modifications, supporting broader future applications.