[Paper Review] First (calibration) experiment using proton beam from FRENA at SINP
This study presents the first calibration of the 3 MV Tandetron accelerator FRENA at SINP using the 7Li(p,n)7Be reaction threshold. By measuring neutron yields across proton energies from 1.78 to 2.5 MeV, the threshold energy was determined as 1872 ± 2.8 keV, indicating a 6–10 keV energy shift, validating the accelerator's terminal voltage calibration for future nuclear astrophysics experiments.
This work presents the first calibration experiment of a 3 MV Tandetron accelerator, FRENA, performed in May 2022. The $^7$Li(p,n) reaction threshold was measured to calibrate the terminal voltage measuring device. A LiF target of thickness 175 $μ$g/cm$^2$ was used in the experiment. The measured threshold was 1872$\pm$2.7 keV, indicating 6$-$10 keV energy shift.
Motivation & Objective
- To calibrate the terminal voltage of the 3 MV Tandetron accelerator FRENA at SINP, essential for accurate energy delivery in low-energy nuclear astrophysics experiments.
- To validate the accelerator's energy scale by measuring the threshold energy of the 7Li(p,n)7Be reaction, a well-known reaction with a precisely known theoretical threshold.
- To establish a reliable calibration procedure using threshold reactions to ensure accurate beam energy measurements before full-scale nuclear astrophysics campaigns.
- To lay the foundation for future calibration using additional (p,n), (α,n), (p,γ), and (α,γ) reactions across the FRENA energy range.
Proposed method
- Conducted the 7Li(p,n)7Be reaction using a 175 μg/cm² LiF target on a 130 μg/cm² Al backing at the PBA beamline of FRENA.
- Irradiated the target with proton beams ranging from 1.78 to 2.5 MeV at ~100 nA beam current, with beam energy measured via terminal voltage.
- Detected emitted neutrons using a cylindrical BF₃ detector (10 mm × 10 cm) housed in a high-density polyethylene moderator, positioned 1 m from the target and normal to the beam.
- Shielded the beam dump with concrete, high-density polyethylene, and lead to suppress background neutrons and γ-rays.
- Acquired pulse height spectra at each proton energy and integrated them to determine neutron flux per beam charge.
- Applied a 2/3rd power transformation to the neutron flux data to linearize the (Eₚ - Eₜₕ)³/² dependence near threshold, enabling precise threshold extraction via linear fitting.
Experimental results
Research questions
- RQ1What is the experimentally measured threshold energy for the 7Li(p,n)7Be reaction on FRENA?
- RQ2How does the measured threshold compare to the known theoretical value of 1.88 MeV, and what does this imply about the accelerator's terminal voltage calibration?
- RQ3To what extent does the neutron yield follow the expected (Eₚ - Eₜₕ)³/² energy dependence near threshold?
- RQ4Can the 7Li(p,n)7Be reaction serve as a reliable calibration standard for the FRENA accelerator's energy scale?
- RQ5What is the magnitude of the energy shift between the measured and expected threshold, and how does it affect accelerator performance?
Key findings
- The measured threshold energy for the 7Li(p,n)7Be reaction was 1872.5 ± 2.8 keV, determined from the intersection of two linear fits to the 2/3rd power of neutron flux versus proton energy.
- The measured threshold lies 8.0 keV below the known theoretical threshold of 1.88 MeV, indicating a 6–10 keV energy shift in the accelerator's terminal voltage calibration.
- Neutron yield data showed a clear (Eₚ - Eₜₕ)³/² dependence above threshold, confirmed by the linear fit of the 2/3rd power of flux versus Eₚ.
- Background contributions from the blank frame were negligible, confirming effective suppression of spurious signals.
- The accelerator beam current was stable at ~100 nA, and the Faraday cup confirmed consistent beam delivery across the energy range.
- The results validate the use of the 7Li(p,n)7Be reaction as a reliable calibration tool for FRENA, with a precision of ±2.8 keV.