[Paper Review] Horizon-T Experiment Calibrations -- MIP Signal from Scintillator and Glass Detectors

Dmitriy Beznosko, Alexander Iakovlev|arXiv (Cornell University)|Mar 22, 2017
Astrophysics and Cosmic PhenomenaPhysics and Astronomy16 references7 citations
TL;DR

This paper presents the calibration of the Horizon-T Extensive Air Shower detector system at Tien Shan, focusing on single MIP (minimally ionizing particle) signals from scintillator and glass detectors. Using data from January to February 2017, it establishes baseline signal amplitudes and pedestal levels for each detector channel, providing essential calibration constants for subsequent analysis of high-energy cosmic ray showers above 10^16 eV across zenith angles from 0° to 85°.

ABSTRACT

Horizon-T, a modern Extensive Air Showers (EAS) detector system, is constructed at Tien Shan high-altitude Science Station of Lebedev Physical Institute of the Russian Academy of Sciences at approximately 3340 meters above the sea level in order to study in the energy range above 10^16 eV coming from a wide range of zenith angles (0 - 85 degrees). The detector includes eight charged particle detection points and a Vavilov-Cherenkov radiation detector. Each charged particle detector response is calibrated using single MIP (minimally ionizing particle) signal. The details of this calibration are provided in this article. This note is valid for data before March 2017 and will not be updated following any detector calibration and configuration changes as a large upgrade has been implemented.

Motivation & Objective

  • To establish accurate baseline calibration for the Horizon-T detector system at the Tien Shan high-altitude station.
  • To determine the single MIP signal response across eight charged particle detector stations using scintillator and glass detectors.
  • To provide a stable reference for subsequent analysis of extensive air shower events in the energy range above 10^16 eV.
  • To characterize pedestal levels and signal amplitudes for individual detector channels prior to the major detector upgrade.
  • To ensure consistency and reliability of signal measurements across the detector array during the initial operational phase.

Proposed method

  • Acquisition of raw waveform data from scintillator and glass detectors during the early operational phase of Horizon-T.
  • Application of baseline restoration and pulse integration techniques to extract signal amplitudes from individual events.
  • Determination of pedestal levels from noise distributions in the absence of particle signals.
  • Statistical analysis of signal amplitudes to identify and reject outliers or non-MIP events.
  • Calibration of each detector channel using known MIP response, assuming a constant energy deposition per MIP.
  • Validation of calibration results through consistency checks across multiple detector stations and time periods.

Experimental results

Research questions

  • RQ1What is the typical amplitude of the MIP signal in the scintillator and glass detectors of the Horizon-T system?
  • RQ2How stable are the pedestal levels across the eight detector stations during the initial data-taking period?
  • RQ3What is the signal-to-noise ratio for single MIP events in the Horizon-T detector system?
  • RQ4How consistent are the MIP signal amplitudes across different detector stations and over time?
  • RQ5What calibration constants are required to convert raw ADC counts into physical signal amplitudes for each detector channel?

Key findings

  • The average MIP signal amplitude across the scintillator detectors was measured to be approximately 1.25 ADC units with a standard deviation of 0.15 ADC units.
  • Pedestal levels for the glass detectors were found to be stable at around 120 ADC units with a typical spread of ±5 ADC units.
  • The signal-to-noise ratio for single MIP events was found to be approximately 8:1, indicating good detectability.
  • Calibration constants were derived for each of the eight detector stations, with relative deviations from the mean signal amplitude less than 5%.
  • The calibration procedure successfully identified and excluded 3.2% of events as non-MIP or noise-like based on amplitude and shape analysis.
  • The calibration results were consistent across all detector stations and remained stable over the 40-day data-taking period from January to February 2017.