[Paper Review] Glass-based charged particle detector performance for Horizon-T EAS detector system
This paper presents a glass-based charged particle detector designed for the Horizon-T extensive air shower (EAS) detector system at Tien Shan, utilizing fast-response scintillating glass to improve ultra-high energy cosmic ray detection resolution. The detector achieves stable pulse responses and validated calibration through simulation and comparison with scintillator data, demonstrating reliable single-particle response and wide dynamic range for high-precision EAS measurements.
An implementation of a novel of glass-based detector with fast response and wide detection range is needed to increase resolution for ultra-high energy cosmic rays detection. Such detector has been designed and built for the Horizon-T detector system at Tien Shan high-altitude Science Station. The main characteristics, such as design, duration of the detector pulse and calibration of a single particle response are discussed.
Motivation & Objective
- To enhance the angular and energy resolution of ultra-high energy cosmic ray (UHECR) detection at the Horizon-T EAS detector system.
- To develop a glass-based charged particle detector with fast response and broad dynamic range for improved EAS shower reconstruction.
- To calibrate the single-particle response of the glass detector using simulation validated against experimental scintillator measurements.
- To assess detector performance through pulse duration and signal response characterization for integration into the Horizon-T array.
Proposed method
- Design and construction of a glass-based charged particle detector using fast-response scintillating glass as the active material.
- Use of Monte Carlo simulation to model detector response and validate performance against experimental data from scintillator detectors.
- Pulse shape analysis to determine the duration and amplitude of signals induced by single charged particles.
- Calibration of the detector's single-particle response using simulated energy deposition and time response profiles.
- Comparison of simulated glass detector output with measured scintillator detector data to validate the simulation model.
- Final performance assessment summarized in a table, including pulse width, signal amplitude, and dynamic range.
Experimental results
Research questions
- RQ1How does the glass-based detector's pulse response compare to that of conventional scintillator detectors in terms of timing and amplitude?
- RQ2What is the achievable single-particle response resolution and dynamic range of the glass detector in simulated conditions?
- RQ3To what extent does the simulation accurately reproduce experimental scintillator data for validation purposes?
- RQ4How stable and repeatable is the detector's signal output across different particle energies and incident angles?
Key findings
- The glass-based detector achieves a pulse width of approximately 10 ns, indicating fast signal response suitable for high-precision timing in EAS detection.
- The simulation results show good agreement with experimental scintillator measurements, validating the model used for glass detector performance assessment.
- The detector demonstrates a wide dynamic range in signal response, enabling detection of particles across a broad energy spectrum.
- Single-particle response is well-calibrated in simulation, with consistent amplitude and timing characteristics across multiple events.
- The final performance summary table confirms the detector's suitability for integration into the Horizon-T EAS system.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.