[Paper Review] JWST MIRI flight performance: The Medium-Resolution Spectrometer

Ioannis Argyriou, Alistair Glasse|arXiv (Cornell University)|Mar 23, 2023
Calibration and Measurement TechniquesEngineering12 citations
TL;DR

This paper documents the in-flight performance of the JWST MIRI Medium-Resolution Spectrometer (MRS), detailing optical, spectral, and spectro-photometric performance and calibration improvements achieved in flight.

ABSTRACT

The Medium-Resolution Spectrometer (MRS) provides one of the four operating modes of the Mid-Infrared Instrument (MIRI) on board the James Webb Space Telescope (JWST). The MRS is an integral field spectrometer, measuring the spatial and spectral distributions of light across the 5-28 $μm$ wavelength range with a spectral resolving power between 3700-1300. We present the MRS's optical, spectral, and spectro-photometric performance, as achieved in flight, and we report on the effects that limit the instrument's ultimate sensitivity. The MRS flight performance has been quantified using observations of stars, planetary nebulae, and planets in our Solar System. The precision and accuracy of this calibration was checked against celestial calibrators with well-known flux levels and spectral features. We find that the MRS geometric calibration has a distortion solution accuracy relative to the commanded position of 8 mas at 5 $μm$ and 23 mas at 28 $μm$. The wavelength calibration is accurate to within 9 km/sec at 5 $μm$ and 27 km/sec at 28 $μm$. The uncertainty in the absolute spectro-photometric calibration accuracy was estimated at 5.6 +- 0.7 %. The MIRI calibration pipeline is able to suppress the amplitude of spectral fringes to below 1.5 % for both extended and point sources across the entire wavelength range. The MRS point spread function (PSF) is 60 % broader than the diffraction limit along its long axis at 5 $μm$ and is 15 % broader at 28 $μm$. The MRS flight performance is found to be better than prelaunch expectations. The MRS is one of the most subscribed observing modes of JWST and is yielding many high-profile publications. It is currently humanity's most powerful instrument for measuring the mid-infrared spectra of celestial sources and is expected to continue as such for many years to come.

Motivation & Objective

  • Assess the optical fidelity and geometric/spectral distortion of the MRS in flight.
  • Quantify the spectro-photometric precision and absolute calibration accuracy.
  • Identify systematic effects limiting the MRS sensitivity and compare with preflight expectations.
  • Describe the calibration pipeline and fringe correction strategies used in flight.
  • Provide updated wavelength and distortion solutions derived from on-sky observations.

Proposed method

  • Quantify performance using observations of stars, planetary nebulae, and Solar System planets.
  • Derive and apply geometric distortion and wavelength calibration solutions per MRS slice and subband.
  • Evaluate PSF broadening and detector scattering effects, and model/subtract scattered light at the detector plane.
  • Characterize and correct spectral fringing using pipeline fringe flats and residual correction algorithms.
  • Validate spectro-photometric accuracy against standard stars and calibrated sources.
  • Compare flight performance and derived calibrations against preflight expectations.

Experimental results

Research questions

  • RQ1What is the in-flight geometric distortion accuracy of the MRS across all channels and slices?
  • RQ2What is the wavelength calibration accuracy and how does it improve with on-sky data?
  • RQ3What are the dominant systematic effects (e.g., detector scattering, fringing) and how effectively are they mitigated in flight?
  • RQ4How does the MRS perform in terms of spectro-photometric absolute calibration and flux stability?
  • RQ5How does the in-flight PSF and spectral resolving power compare with ground-based models and expectations?

Key findings

  • Geometric distortion accuracy is 8 mas at 5 μm and 23 mas at 28 μm.
  • Wavelength calibration accuracy improves to about 6 km/s with FLT-5 solution across bands.
  • Absolute spectro-photometric calibration accuracy is 5.6 ± 0.7%.
  • Spectral fringes can be suppressed to below 1.5% amplitude across the range using pipeline corrections.
  • Detector internal scattering causes PSF broadening up to 60% along the long axis at 5 μm, decreasing at longer wavelengths; a subtraction algorithm reduces this flux to below 1% impact.
  • MRS PSF is 60% broader than diffraction-limited along the long axis at 5 μm and 15% broader at 28 μm.