Skip to main content
QUICK REVIEW

[Paper Review] Module concept and thermo-mechanical studies of the silicon-based TT-PET small-animal scanner

D. Ferrère, Y. Bandi|arXiv (Cornell University)|Nov 29, 2018
Nuclear Physics and ApplicationsPhysics and Astronomy1 references3 citations
TL;DR

This paper presents a novel module concept and thermo-mechanical design for a compact, high-resolution small-animal TOF PET scanner using monolithic silicon pixel detectors. It demonstrates a stacked-die wire bonding technique for interconnecting 1920 ASICs and validates a micro-channel ceramic cooling system that successfully manages 300 W of heat dissipation, with FEA simulations matching measurements within 1 °C after HTC optimization.

ABSTRACT

The TT-PET collaboration is developing an MRI-compatible small animal PET scanner in which the sensitive element is a monolithic silicon pixel ASIC targeting 30 ps RMS time resolution. The photon-detection technique is based on a stack of alternating layers of high-Z photon converter and 100 $\mathrm{μm}$ silicon sensors, to produce a scanner with 0.5 $\mathrm{ imes}$ 0.5 $\mathrm{ imes}$ 0.2 $\mathrm{mm^{3}}$ granularity for precise depth-of-interaction measurement. In this paper we present the results of simulation studies for the expected data rate, time-of-flight and spatial resolution, as well as the performance of image reconstruction with and without the use of timing information.

Motivation & Objective

  • Develop a compact, high-resolution small-animal PET scanner compatible with MRI systems.
  • Address extreme mechanical, thermal, and electrical constraints in a 5 cm diameter, 3.9 cm inner-diameter cylindrical geometry.
  • Minimize detector dead area while enabling full electrical interconnection of 1.5 million pixels across 1920 ASICs.
  • Design a cooling system capable of dissipating 300 W within tight spatial constraints.
  • Validate the thermo-mechanical performance of a full-scale mock-up using FEA and experimental measurements.

Proposed method

  • Use monolithic silicon pixel timing detectors with 30 ps rms time resolution, bonded via 5 µm double-sided adhesive tape.
  • Implement a stacked-die wire bonding technique for daisy-chaining ASICs across 60 modules per tower.
  • Integrate micro-channel ceramic cooling blocks with water coolant to extract 300 W from the scanner volume.
  • Construct a thermo-mechanical mock-up with 12 super-modules and 12 cooling blocks to test thermal performance.
  • Perform FEA simulations with boundary conditions including convective cooling (HTC = 1 W m⁻² K⁻¹), fixed base temperature (19 °C), and heat fluxes of 2.4 mW mm⁻² on heater pads.
  • Optimize HTC values (5000 W m⁻² K⁻¹ for Al, 9000 W m⁻² K⁻¹ for Al₂O₃) to match simulation with experimental temperature measurements.

Experimental results

Research questions

  • RQ1Can a stacked-die wire bonding technique reliably interconnect 1920 monolithic ASICs in a 3D-stacked, wedge-shaped detector with minimal dead area?
  • RQ2Can micro-channel ceramic cooling blocks effectively dissipate 300 W of heat within a 5 cm diameter, 8.4 cm outer-diameter cylindrical volume?
  • RQ3How accurately can FEA simulations predict the thermal behavior of the scanner mock-up under full-power operation?
  • RQ4What HTC values are required to achieve <1 °C agreement between FEA simulations and experimental temperature measurements?
  • RQ5Can a stress-release mechanism prevent mechanical stress transfer to wire bonds during radial interconnection of super-modules?

Key findings

  • The stacked-die wire bonding technique was successfully tested on dummy substrates, proving reliable and reproducible for daisy-chaining ASICs.
  • The thermo-mechanical mock-up achieved a maximum temperature difference of 2 °C between FEA simulation and measurement with initial HTC assumptions.
  • After parametric optimization, FEA simulations matched experimental temperatures with an average error of less than 1 °C across 12 modules.
  • The optimal HTC values were determined to be 5000 W m⁻² K⁻¹ for the aluminum cooling block and 9000 W m⁻² K⁻¹ for the Al₂O₃ block.
  • Temperature mapping showed a 2–3 °C discrepancy between cooling blocks due to asymmetric heat transfer, which was accurately captured in the FEA model.
  • The mock-up demonstrated that the cooling system can handle 36 W of total power (2.4 mW mm⁻²) with coolant inlet and outlet temperatures of 11.8 °C and 16 °C, respectively.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.