Skip to main content
QUICK REVIEW

[Paper Review] MPS and ACS with an atomic magnetometer

Simone Colombo, Victor Lebedev|arXiv (Cornell University)|Nov 27, 2015
Characterization and Applications of Magnetic NanoparticlesEngineering6 citations
TL;DR

This paper demonstrates that a single atomic magnetometer operating in an unshielded environment can perform high-sensitivity magnetic particle spectroscopy (MPS) and AC susceptometry (ACS) on liquid-suspended magnetic nanoparticles. By detecting the derivative of magnetization (dM/dH) via frequency-modulation spectroscopy, the method achieves sub-microgram iron detection limits and enables direct, quantitative measurement of M(H) and dM/dH at low frequencies (≤1 Hz), paving the way for low-frequency, high-sensitivity X-space MPI scanners using atomic magnetometers instead of conventional induction coils.

ABSTRACT

We show that a single atomic magnetometer in a magnetically unshielded environment can be used to perform magnetic particle spectroscopy (MPS) and AC susceptometry (ACS) on liquid-suspended magnetic nanoparticles. We demonstrate methods allowing a simultaneous recording of M(H) and dM/dH(H) dependences of samples containing down to 1 $μ$g of iron. Our results pave the way towards an atomic magnetometer based MPI scanner.

Motivation & Objective

  • To develop a low-frequency, high-sensitivity alternative to conventional MPI using atomic magnetometers instead of induction coils.
  • To enable quantitative measurement of magnetization (M(H)) and its derivative (dM/dH) in unshielded environments.
  • To demonstrate sub-µg iron detection sensitivity for magnetic nanoparticle suspensions.
  • To explore the feasibility of using atomic magnetometers as the core sensor in a future X-space MPI scanner.
  • To overcome limitations of conventional MPI, such as high-frequency drive field requirements and SAR concerns in biomedical applications.

Proposed method

  • An optically pumped atomic magnetometer (OPM) detects magnetic field changes from magnetic nanoparticles (MNPs) via Larmor precession of spin-polarized cesium atoms in a 27 µT bias field.
  • The OPM operates in a phase-locked loop (PLL) configuration to maintain resonance, encoding magnetic field changes into frequency shifts (δf NP ∝ δB⋅B̂₀).
  • A lock-in amplifier extracts the signal, and sideband demodulation enables simultaneous acquisition of M(H) and dM/dH(H) from the same measurement.
  • The method uses frequency-modulation (FM) spectroscopy principles to isolate the dM/dH response, which is insensitive to low-frequency drifts in the bias field.
  • The system applies a slow sinusoidal scan field (f_scan = 600 mHz) to the MNP sample, and the OPM records the resulting time-dependent field modulation.
  • Fourier transformation of the OPM signal allows extraction of both the magnetization curve M(H) and the derivative dM/dH, with calibration via δf NP = γF δB⋅B̂₀.

Experimental results

Research questions

  • RQ1Can an atomic magnetometer in an unshielded environment perform quantitative MPS and ACS on magnetic nanoparticles?
  • RQ2What is the minimum iron content detectable using this OPM-based MPS/ACS method?
  • RQ3How does the FM-spectroscopy approach improve the stability and signal-to-noise ratio of dM/dH measurements compared to direct M(H) recording?
  • RQ4Can the OPM-based method detect the first harmonic response of MNPs, which is typically lost in conventional MPI?
  • RQ5Is it feasible to use atomic magnetometers as the primary sensor in a future X-space MPI scanner?

Key findings

  • The method achieves a detection limit of approximately 700 ng of iron in a 50-second recording, corresponding to a sub-µg iron sensitivity.
  • The OPM simultaneously records both M(H) and dM/dH(H) with high fidelity, enabling direct quantitative analysis of magnetization and its derivative.
  • The FM-spectroscopy-based dM/dH measurement is robust against low-frequency noise and drift in the bias field, unlike direct M(H) recording.
  • The Fourier spectrum of dM/dH shows dominant odd harmonics, while the M(H) spectrum is affected by low-frequency noise, confirming the superior stability of the derivative measurement.
  • The method successfully measures the magnetization response of large, hydrodynamically blocked particles at low frequencies (≤1 Hz), which are inaccessible to standard 25 kHz MPI systems.
  • The results demonstrate the feasibility of using atomic magnetometers as a low-frequency, high-sensitivity alternative to conventional induction coils in X-space MPI.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.