Skip to main content
QUICK REVIEW

[Paper Review] Recognising Cardiac Abnormalities in Wearable Device Photoplethysmography (PPG) with Deep Learning

Stewart Whiting, Samuel Moreland|arXiv (Cornell University)|Jul 11, 2018
Non-Invasive Vital Sign MonitoringEngineering8 references6 citations
TL;DR

This paper proposes a deep learning approach using an LSTM network to detect cardiac abnormalities—particularly premature ventricular contractions (PVCs)—directly from wearable photoplethysmography (PPG) signals, achieving 60% sensitivity and 77% specificity in identifying ECG-confirmed PVCs without requiring ECG data, demonstrating the feasibility of PPG-only cardiac monitoring in real-world settings.

ABSTRACT

Cardiac abnormalities affecting heart rate and rhythm are commonly observed in both healthy and acutely unwell people. Although many of these are benign, they can sometimes indicate a serious health risk. ECG monitors are typically used to detect these events in electrical heart activity, however they are impractical for continuous long-term use. In contrast, current-generation wearables with optical photoplethysmography (PPG) have gained popularity with their low-cost, lack of wires and tiny size. Many cardiac abnormalities such as ectopic beats and AF can manifest as both obvious and subtle anomalies in a PPG waveform as they disrupt blood flow. We propose an automatic method for recognising these anomalies in PPG signal alone, without the need for ECG. We train an LSTM deep neural network on 400,000 clean PPG samples to learn typical PPG morphology and rhythm, and flag PPG signal diverging from this as cardiac abnormalities. We compare the cardiac abnormalities our approach recognises with the ectopic beats recorded by a bedside ECG monitor for 29 patients over 47.6 hours of gold standard observations. Our proposed cardiac abnormality recognition approach recognises 60%+ of ECG-detected PVCs in PPG signal, with a false positive rate of 23% - demonstrating the compelling power and value of this novel approach. Finally we examine how cardiac abnormalities manifest in PPG signal for in- and out-of-hospital patient populations using a wearable device during standard care.

Motivation & Objective

  • To develop an automated method for detecting cardiac abnormalities in PPG signals without relying on ECG.
  • To evaluate the performance of a deep learning model in recognizing PVCs from wearable PPG data against a clinical ECG gold standard.
  • To analyze how cardiac abnormalities manifest in PPG signals across diverse patient populations in both in-hospital and outpatient settings.
  • To explore the potential of PPG-only monitoring for improving long-term cardiac risk detection and early warning systems.

Proposed method

  • An LSTM deep neural network is trained on 400,000 clean PPG samples to learn normal PPG morphology and rhythm.
  • The model flags PPG signals that deviate significantly from the learned normal pattern as potential cardiac abnormalities.
  • A comparative evaluation is performed using 29 patients monitored with both wearable PPG and bedside ECG for 47.6 hours, with PVCs serving as the gold standard.
  • The method includes high-motion PPG signal filtering to ensure signal quality and alignment with ECG observations.
  • Performance is evaluated using a set-based confusion matrix, comparing PPG-detected anomalies with ECG-confirmed PVCs.
  • Additional analysis aggregates PPG-based abnormality frequency across multiple clinical populations (HDU, AMU, ED, SURGERY, HOME).

Experimental results

Research questions

  • RQ1Can a deep learning model trained on normal PPG signals detect cardiac abnormalities such as PVCs without ECG reference?
  • RQ2How does the performance of PPG-only detection compare to ECG gold standard in real-world ambulatory monitoring?
  • RQ3How do the frequencies and patterns of cardiac abnormalities in PPG signals vary across different patient populations?
  • RQ4What is the impact of signal quality and motion artifacts on the detection accuracy of PPG-based cardiac anomaly detection?

Key findings

  • The proposed method detects 60% of ECG-confirmed PVCs in PPG signals when at least one PVC per minute is present in the gold standard.
  • The false positive rate is 23%, indicating that about 23% of PPG signals flagged as abnormal do not contain PVCs according to ECG.
  • When the gold standard includes two or more PVCs per minute, detection sensitivity increases to 68%, showing improved performance with higher PVC burden.
  • In-hospital populations such as HDU and AMU show higher PPG-based abnormality rates (5.1% and 5.8% respectively), reflecting higher patient acuity.
  • Surgically treated patients (SURGERY) show the lowest abnormality rate (1.54%), possibly due to better health status or clinical monitoring.
  • The ED population shows high variability in abnormality frequency (mean 3.17%, standard deviation 7.9%), indicating heterogeneous patient acuity.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.