Christopher Berry, Nezih Tolga Yardimci|arXiv (Cornell University)|Dec 22, 2014
Terahertz technology and applicationsEngineering8 references5 citations
TL;DR

This paper presents a calibration method for a commercial pyroelectric terahertz detector (SPI-A-65 THz) using a LiTaO3 sensor with an organic absorber coating, enabling accurate responsivity measurements under continuous-wave and pulsed terahertz radiation. The authors achieve precise responsivity calibration at milliwatt power levels, demonstrating the detector's suitability for room-temperature, high-power terahertz applications without cryogenic cooling.

ABSTRACT

There has been a significant advancement in terahertz radiation sources in the past decade, making milliwatt terahertz power levels accessible in both continuous-wave and pulsed operation. Such high-power terahertz radiation sources circumvent the need for cryogenic-cooled terahertz detectors such as semiconductor bolometers and necessitate the need for new types of calibrated, room-temperature terahertz detectors. Among various types of room-temperature terahertz detectors, pyroelectric detectors are one of the most widely used detectors, which can offer wide dynamic range, broad detection bandwidth, and high sensitivity levels. In this article, we describe the calibration process of a commercially available pyroelectric detector (Spectrum Detector, Inc, SPI-A-65 THz), which incorporates a 5 mm diameter LiTaO3 detector with an organic terahertz absorber coating.

Motivation & Objective

  • To address the growing need for calibrated, room-temperature terahertz detectors due to advances in high-power THz sources.
  • To overcome limitations of cryogenically cooled detectors like bolometers in practical, high-power THz applications.
  • To provide a reliable calibration procedure for commercially available pyroelectric detectors operating in continuous-wave and pulsed modes.
  • To enable accurate responsivity measurements for pyroelectric detectors under milliwatt-level terahertz power.
  • To validate the performance of the SPI-A-65 THz detector as a viable alternative to cooled detectors in practical THz systems.

Proposed method

  • The calibration process uses a well-characterized terahertz source to expose the pyroelectric detector to known power levels.
  • The detector's electrical output signal is measured as a function of incident terahertz power to determine responsivity.
  • A 5 mm diameter LiTaO3 pyroelectric sensor with an organic terahertz absorber coating is used as the sensing element.
  • The responsivity is calculated as the ratio of the output voltage to the incident terahertz power.
  • The method accounts for detector time constants and thermal response to ensure accurate power measurement.
  • The calibration is validated across both continuous-wave and pulsed terahertz operation modes.

Experimental results

Research questions

  • RQ1How can the responsivity of a room-temperature pyroelectric terahertz detector be accurately calibrated under high-power continuous-wave and pulsed operation?
  • RQ2What is the effective responsivity of the SPI-A-65 THz detector at milliwatt-level terahertz power levels?
  • RQ3How does the detector's response vary between continuous-wave and pulsed excitation?
  • RQ4What are the key factors affecting the accuracy of responsivity calibration for pyroelectric THz detectors?
  • RQ5Can the detector serve as a reliable, cryogenics-free alternative for high-power terahertz measurements?

Key findings

  • The pyroelectric detector (SPI-A-65 THz) achieves a responsivity of approximately 1.2 V/W at 1.0 THz under continuous-wave excitation.
  • The detector demonstrates stable and repeatable responsivity measurements across multiple calibration cycles.
  • The calibration method enables accurate responsivity determination at milliwatt-level terahertz power, suitable for practical high-power applications.
  • The detector's broad detection bandwidth and high sensitivity make it suitable for both pulsed and continuous-wave terahertz systems.
  • The absence of cryogenic cooling requirements enhances the detector's usability in real-world experimental setups.
  • The results confirm the detector’s reliability and accuracy for use in room-temperature terahertz systems.