[Paper Review] Searching for magnetic monopoles trapped in accelerator material at the Large Hadron Collider
This study uses a SQUID-based magnetometer to search for magnetic monopoles trapped in obsolete LHC beam pipe material exposed to proton-proton and lead-lead collisions. Despite testing multiple calibration methods and assessing background signals, no monopoles were detected with magnetic charge above approximately one-third of the Dirac charge, establishing a lower detection threshold for future searches.
If produced in high energy particle collisions at the LHC, magnetic monopoles could stop in material surrounding the interaction points. Obsolete parts of the beam pipe near the CMS interaction region, which were exposed to the products of pp and heavy ion collisions, were analysed using a SQUID-based magnetometer. The purpose of this work is to quantify the performance of the magnetometer in the context of a monopole search using a small set of samples of accelerator material ahead of the 2013 shutdown.
Motivation & Objective
- To develop and test experimental techniques for detecting magnetic monopoles trapped in accelerator materials at the LHC.
- To quantify the sensitivity of a SQUID-based magnetometer to monopoles with varying magnetic charges.
- To identify and mitigate false signals from large magnetic dipoles in sample materials.
- To establish protocols for handling and releasing accelerator material for monopole searches post-2013 LHC shutdown.
- To demonstrate the feasibility of detecting monopoles with charges both below and above the Dirac charge.
Proposed method
- A SQUID-based magnetometer was used to measure persistent currents induced by magnetic charges in beam pipe samples.
- Calibration was performed using pseudopoles with known magnetic charges, including a 4.14g_D standard, to characterize instrument response.
- Magnetic field simulations and rock samples with large dipole moments were used to study background signals resembling monopole signatures.
- Samples were measured multiple times at 48 or more positions to ensure reproducibility and detect offset jumps.
- Demagnetisation with oscillating fields was applied to reduce background magnetisation in selected samples.
- Background subtraction was performed using empty holder measurements taken between sample runs.
Experimental results
Research questions
- RQ1What is the minimum detectable magnetic charge using a SQUID-based magnetometer in LHC accelerator material?
- RQ2How do large magnetic dipole moments in material samples mimic monopole signals, and can they be distinguished?
- RQ3What is the expected response of the magnetometer to monopoles with charges ranging from below to above the Dirac charge?
- RQ4Can monopoles with charges greater than one-third of the Dirac charge be detected in real accelerator samples?
- RQ5What protocols are necessary to safely extract and analyze accelerator material for monopole searches?
Key findings
- No evidence was found for magnetic monopoles with charges greater than approximately one-third of the Dirac charge in the tested beam pipe samples.
- Peak induced currents in the samples ranged from 0 to 0.35 times the current expected for a Dirac monopole, indicating no significant monopole signal.
- Offset jumps of ~0.15g_D were observed, likely due to remnant magnetisation, and were reduced by demagnetisation.
- The magnetometer response was successfully calibrated using a pseudopole with charge 4.14g_D, confirming detectability of such signals.
- The persistent current signature from a trapped monopole was shown to be distinct and reproducible, matching the expected pattern.
- Background signals from magnetic dipoles were identified and quantified, demonstrating the need for careful sample selection and demagnetisation.