[Paper Review] Skin color independent robust assessment of capillary refill time
This study proposes a skin color-independent, robust method for capillary refill time (pCRT) assessment using an RGB camera, LED lighting, and crossed circular polarizers to reduce surface reflection. By applying 7 kPa pressure for 5 seconds and analyzing green channel intensity decay via exponential regression with a data-dependent cut-off and uncertainty-based flagging, the method achieves high repeatability (80% of measurements within ±20% of individual mean) across all Fitzpatrick skin phototypes I–VI.
Capillary refill time (CRT) is a method for evaluating peripheral perfusion by visual assessment. CRT is especially useful for quick evaluations in the absence of sophisticated equipment. However, there are repeatability and reproducibility limitations with CRT, especially for dark skin. To test the limits of CRT repeatability and skin color independence, we developed a system and method to perform simple and robust CRT measurements. The system consists of an RGB camera and an LED lamp, with crossed circular polarizers imaging to attenuate the light reflected by the superficial layer of the skin. The capillary refill time is determined using an exponential regression on the time-dependent green channel mean pixel intensity of the region of interest after the compression is released. We limited this regression up to a data-dependent cut-off time, after which we assume the exponential model is invalid, and used the confidence interval of the uncertainty to develop a criterion to flag and discard faulty measurements. We tested the system on twenty-two volunteers with skin phototypes ranging from I to VI on the Fitzpatrick scale, applying to their forearms a 7 kPa compression for 5 s. After the release of measurements flagged as inadequate (about 20\% of measurements) by our regressions, our results indicated good precision, with high repeatability for all skin phototypes. Approximately 80\% of measurements fall within $\pm 20\%$ of the individual's expected value for CRT (mean CRT value). Our results suggest CRT can be used as a quantitative measurement and encourages further developments for the implementation of a similar method on smartphone cameras for quick and robust CRT measurements in patients' triage, monitoring, and telehealth.
Motivation & Objective
- Address the lack of standardization and poor reproducibility in manual capillary refill time (CRT) measurements, especially for dark skin.
- Overcome limitations in visual CRT assessment due to observer variability, lighting, temperature, and skin phototype effects.
- Develop a low-cost, robust, and objective method for quantitative CRT measurement that is independent of skin color and observer bias.
- Ensure measurement reliability by flagging and discarding low-quality readings using regression uncertainty and a dynamic cut-off time.
- Enable future implementation on smartphone cameras for point-of-care, telehealth, and continuous monitoring applications.
Proposed method
- Use an RGB camera and LED light source with crossed circular polarizers to minimize surface reflection and enhance signal-to-noise ratio in skin imaging.
- Apply a controlled 7 kPa compression for 5 seconds on the forearm using a custom cylindrical device with a Teflon contact surface.
- Record a video of the region of interest (ROI) after compression release and extract time-series mean green channel intensity.
- Fit an exponential decay model to the intensity recovery curve using nonlinear regression, with a data-dependent cut-off time to exclude invalid post-refill phases.
- Use the confidence interval of the regression to identify and flag unreliable measurements for rejection.
- Define pCRT as the time constant of the exponential fit, representing the capillary refill time under controlled, objective conditions.

Experimental results
Research questions
- RQ1Can capillary refill time be measured reliably and reproducibly across all Fitzpatrick skin phototypes using a low-cost, camera-based system?
- RQ2Does the use of crossed circular polarizers significantly improve measurement robustness by reducing surface reflection artifacts?
- RQ3Can a data-driven cut-off time and regression uncertainty criterion effectively identify and discard faulty measurements?
- RQ4How does low-pressure compression (7 kPa) affect measurement repeatability and accuracy compared to higher pressures used in conventional CRT?
- RQ5To what extent is the proposed method independent of observer bias and skin color, enabling use in diverse clinical populations?
Key findings
- The method achieved high repeatability, with approximately 80% of pCRT measurements falling within ±20% of each individual’s mean CRT value across all skin phototypes.
- About 20% of measurements were flagged and discarded due to poor regression fit or high uncertainty, indicating effective quality control.
- The system demonstrated robustness across all Fitzpatrick skin types (I–VI), confirming skin color independence of the pCRT measurement.
- The use of a low 7 kPa compression pressure improved measurement stability and reduced noise compared to higher pressures, which degraded fit quality.
- The exponential regression model with a dynamic cut-off time significantly improved the reliability and repeatability of pCRT estimation.
- The results support the feasibility of implementing a similar pCRT system on smartphone cameras for non-invasive, objective, and accessible peripheral perfusion monitoring.

Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.