A compilation of Cycle 30 calibration programs for the HST STIS CCD and MAMA detectors, detailing objectives, observations, analyses, and results across multiple monitors (dark, bias, flat fields, CTI/CTE, slit repeatability, dispersion solutions, and sensitivity).
We discuss the suite of STIS calibration programs executed during HST Cycle 30, covering the period 2022 Nov 07 through 2023 Nov 05. For each of the 19 current regular calibration programs, we provide brief descriptions of the objectives, observations, analysis procedures, and results - with comparisons to the results from previous cycles and to desired accuracies, as well as references to more detailed analyses of the calibration data. Many of these calibration programs produce routine reference file deliveries or demonstrate the continuing applicability of existing reference files for processing STIS observations. This ISR provides a brief snapshot of the current instrument performance, similar to those given in annual reports for Cycles 7-10 and 17-21. Two Appendices briefly discuss the state of the onboard calibration lamps and the ongoing major effort to revise the flux calibration for the many STIS spectroscopic and imaging modes.
Motivation & Objective
- Monitor overall baseline performance of the STIS CCD across aging and radiation damage.
- Track dark current, bias structure, read noise, and hot pixels to support data reduction and reference files.
- Characterize charge transfer inefficiency (CTI/CTE) and spurious charge over time.
- Assess spectroscopic and imaging flat fields, focusing, and focus stability for calibration accuracy.
- Monitor wavelength calibration, sensitivity, and dispersion solutions for CCD and MAMA across cycles.
Proposed method
- Full-frame bias and flatfield exposures to measure read noise, CTE, spurious charge, and gain across multiple gains and apertures.
- Darks/superdarks collected daily to track dark current and hot/bad pixels, delivered as reference files.
- Bias and read noise monitoring with weekly superbias files and comparison to previous cycles.
- Hot pixel annealing sessions with pre/post measurements to assess fixed/persisting hot pixels and dark current changes.
- Spectroscopic and imaging flat-field monitoring using various gratings, apertures, and lamp configurations to detect changes and generate p-flats if needed.
- CTI measurements using internal sparse-field tests with cross-dispersed tungsten lamp exposures across multiple y-positions and flux levels.

Experimental results
Research questions
- RQ1How does the STIS CCD performance evolve over Cycle 30 in terms of CTI/CTE, read noise, bias stability, and dark current?
- RQ2Are there detectable changes in flat-field, spectroscopic flat fields, and PSF/throughput that require updated calibration files?
- RQ3What is the repeatability and stability of the slit wheel positions and dispersion solutions over Cycle 30?
- RQ4How do CTI and spurious charge affect faint-source spectroscopy and what corrections are necessary?
- RQ5What are the time-dependent sensitivity trends for CCD and MAMA detectors across different modes?
Key findings
- CTI continues to rise slowly for both parallel and serial transfers; current CTI values measured by EPER are ~0.073% (parallel) and ~0.0024% (serial).
- Spurious charge in central bias rows is ~1.8 e-/pix (gain=1) and ~8.3 e-/pix (gain=4) with lower values near E1 pseudo-apertures; overall trends align with radiation damage progression.
- Dark rates remained relatively stable during Cycle 30 with an average ~0.13 cts/s/pix.
- Hot pixel annealing yields a net reduction in hot pixels on average, though persistent hot pixels remain; comparison with prior cycles shows mixed results on fixed pixel growth.
- Cycle-30 spectroscopic flat-field and imaging flat-field monitoring indicate no significant evolution in the major flat-field features; p-flats for imaging are not immediately required.
- Wavecal zero-point offsets for CCD and MAMA first-order settings generally remain below 0.3 px at nominal central locations; some offsets at E1 suggest potential CCD format rotation effects.