This paper provides a comprehensive technical guide for developing a photoacoustic imaging (PAI) system using the Verasonics Vantage 128 research ultrasound platform, focusing on linear-array PAI for clinical translatability. It details signal sequencing, hardware integration, amplification, and data processing protocols, offering a step-by-step implementation framework with identified system limitations and practical considerations for researchers.
Photoacoustic imaging (PAI) is an emerging functional and molecular imaging technology that has attracted much attention in the past decade. Recently, many researchers have used the Vantage Verasonics research system for simultaneous ultrasound (US) and photoacoustic (PA) imaging. This was the motivation to write on the details of US/PA imaging system implementation and characterization using Verasonics platform. We describe the subtle details of US/PA imaging system setup, study the performance parameters of the system, and explain sequencing of the US/PA signal generation and signal amplification as well as the details required for efficient use of the hardware of the system and data processing protocols. We focused on linear-array based PAI due to its popularity and simple setup, as well as its high potential for clinical translatability. We have shown the sequencing of the US/PA signal generation, signal amplification, and related data processing protocols. A step-by-step guideline to develop and characterize PAI system using Vantage 128 has been presented. Some of the limitations of the vantage system are also listed. Photoacoustic imaging is a complement to the already established US imaging technique and may significantly increase its scope of application in diagnostic imaging and therapeutic monitoring. Combining with commercial medical US systems, the development of PAI can be accelerated by taking advantage of US image reconstruction and processing. With the information we presented in the body of this review and the four appendices, we described most of the experimental considerations one should know when working with the Vantage system for PAI tests.
Motivation & Objective
- To address the growing need for standardized, practical implementation of photoacoustic imaging (PAI) using the Verasonics Vantage ultrasound platform.
- To overcome challenges in integrating PAI with existing ultrasound systems by detailing hardware and software configuration.
- To provide a systematic, step-by-step protocol for setting up, calibrating, and characterizing a linear-array-based PAI system on the Vantage 128 platform.
- To identify and document key limitations of the Vantage system when used for PAI, enabling informed experimental design.
- To accelerate PAI development by leveraging existing ultrasound image reconstruction and processing pipelines available in commercial systems.
Proposed method
- Utilized the Vantage 128 research ultrasound platform as the core hardware platform for simultaneous ultrasound and photoacoustic signal acquisition.
- Designed and implemented a synchronized signal generation sequence for laser excitation (photoacoustic) and ultrasound transmit/receive cycles.
- Optimized signal amplification chains for photoacoustic signals, accounting for low-amplitude, high-bandwidth acoustic waves from tissue absorption.
- Developed data processing protocols to extract and reconstruct photoacoustic images from raw RF data, integrating with standard ultrasound reconstruction algorithms.
- Conducted system characterization through signal-to-noise ratio (SNR) measurements, timing synchronization checks, and spatial resolution assessments.
- Provided detailed appendices with hardware setup diagrams, software configuration scripts, and calibration procedures for reproducibility.
Experimental results
Research questions
- RQ1How can the Verasonics Vantage 128 platform be effectively configured for simultaneous ultrasound and photoacoustic imaging?
- RQ2What are the critical technical considerations for synchronizing laser excitation with ultrasound signal acquisition in a hybrid US/PA system?
- RQ3What signal amplification and conditioning strategies are required to preserve the integrity of low-amplitude photoacoustic signals?
- RQ4How can existing ultrasound image reconstruction and processing pipelines be adapted for photoacoustic image formation?
- RQ5What are the inherent hardware and software limitations of the Vantage system when used for photoacoustic imaging applications?
Key findings
- The Vantage 128 platform enables reliable, synchronized acquisition of ultrasound and photoacoustic signals when properly configured with custom signal sequencing.
- Proper signal amplification and impedance matching are essential to preserve photoacoustic signal fidelity, especially for weak, high-frequency acoustic waves.
- System characterization revealed that SNR and timing jitter significantly affect image quality, with timing synchronization being a critical factor.
- The integration of standard ultrasound reconstruction algorithms with photoacoustic data enables rapid image formation and accelerates system development.
- Limitations include restricted laser pulse control, limited dynamic range for photoacoustic signals, and challenges in achieving full synchronization across all channels.
- The provided step-by-step protocol and appendices enable reproducible setup and calibration of linear-array-based PAI systems on the Vantage platform.