Skip to main content
QUICK REVIEW

[Paper Review] Temperature-insensitive tunable and stable Fabry-Perot cavity for atomic physics

Joshua Ruelle, Martin Hauden|arXiv (Cornell University)|Mar 12, 2026
Advanced Frequency and Time StandardsPhysics and Astronomy0 citations
TL;DR

The paper introduces a piezoelectrically tunable Fabry-Perot cavity with a composite design that cancels thermal expansion near 4.9 K, achieving ~4×10^-13 fractional frequency instability at 1 s without external stabilization.

ABSTRACT

Optical Fabry-Perot cavities are crucial tools for metrology experiments, where they achieve extreme length stability, and for some atomic physics experiments, where tunability to atomic transitions enables atom-light interactions. However, achieving both frequency stability and tunability in a single cavity has remained a challenge, forcing metrology experiments exploiting atom-cavity interactions to rely on external active feedback systems to stabilize the length of the cavity. Here, we describe a piezoelectrically-tunable cavity with a cancellation of the coefficient of thermal expansion at around $5^\circ\mathrm{C}$, achieving fractional frequency instabilities at the $4 imes 10^{-13}$ level for 1~s integration time. This advance eliminates the need for external stabilization in many atom-cavity experiments, making this design ideal for applications such as ultra-stable superradiant lasers and other cavity quantum electrodynamics experiments.

Motivation & Objective

  • Motivate the need for a cavity that is both tunable and highly frequency-stable for atom-light interactions and cQED experiments.
  • Present a composite cavity design that suppresses thermal expansion around a zero-crossing temperature.
  • Demonstrate passive stability performance suitable for applications like ultra-stable superradiant lasers and cavity QED.
  • Characterize optical, thermal, and electrical aspects to identify primary noise contributors and optimization routes.

Proposed method

  • Design of a 50 mm Zerodur spacer with enlarged apertures and corner cuts optimized by finite-element analysis to minimize holding-force fluctuations.
  • Implementation of a tunable cavity using two 2 mm thick PZT transducers with a ~3.3 μm free stroke at 200 V to span several free spectral ranges.
  • Integration of Kovar washers to tailor the thermal expansion coupling between PZT and mirrors, aiming for partial compensation of PZT- and spacer-induced length changes.
  • Passive and active temperature regulation with a tetrahedral mount, Viton balls, and multi-stage Peltier cooling to achieve 0.1 K/√s in-loop temperature stability and low out-of-loop fluctuations.
  • Pound-Drever-Hall locking of a 578 nm laser to the cavity, followed by optical frequency comb comparison to an H-maser to assess fractional frequency stability.
  • Characterization of cavity performance via ring-down finesse measurements and analysis of PZT noise, including filtering strategies to suppress voltage noise and resonances.

Experimental results

Research questions

  • RQ1Can a Fabry-Perot cavity be both tunable and passively stable enough for precision atom-cavity experiments without external stabilization?
  • RQ2What composite-material strategy and temperature operating point enable cancellation of the effective thermal expansion of the cavity?
  • RQ3What are the dominant noise mechanisms limiting short-term cavity frequency stability, and how can they be mitigated?

Key findings

  • Finesse measured at 578 nm was 6920±40 after bake-out, down from 21,000±7,000 prior to bake-out, suggesting contamination during bake-out.
  • Best measured fractional frequency stability is 4×10^-13 at 1 s integration time when PZTs are either disconnected or filtered, at the zero-crossing temperature T0=4.9±0.5 K.
  • Temperature stepping experiments show a zero-crossing of the total coefficient of thermal expansion at T0 near 4.9 K, with αtot(T) ≈ a_tot(T−T0) and a_tot=4.8±0.9×10^-8 K^-2.
  • Voltage-noise filtering of PZTs reduces the flicker frequency noise contribution to the cavity instability to ~1×10^-13.
  • Under realistic operating conditions with the atomic oven on, the cavity maintains ~1×10^-13 stability at 1 s when operated at T0, degrading about a factor of 2 at room temperature with oven gradients.
  • Overall, the cavity demonstrates passive stability at the mid-10^-13 level, with potential to reach ~4×10^-18 for SR-laser performance if atomic-oven-related drifts are mitigated.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.