[Paper Review] The Fibre Multi-Object Spectrograph (FMOS) for Subaru Telescope

Masahiko Kimura, Toshinori Maihara|arXiv (Cornell University)|Jun 15, 2010
Astronomy and Astrophysical ResearchPhysics and Astronomy7 citations
TL;DR

FMOS is a near-infrared, fiber-fed multi-object spectrograph for the Subaru Telescope, enabling simultaneous spectroscopy of up to 400 objects within a 30′ diameter field of view. It achieves this via a novel fiber positioner (Echidna), a wide-field corrector, and two cooled spectrographs, delivering a system efficiency of 10% and limiting magnitudes of J = 20.1 mag and H = 19.8 AB mag for S/N = 5 in 1-hour integrations.

ABSTRACT

Fibre Multi-Object Spectrograph (FMOS) is the first near-infrared instrument with a wide field of view capable of acquiring spectra simultaneously from up to 400 objects. It has been developed as a common-use instrument for the F/2 prime-focus of the Subaru Telescope. The field coverage of 30' diameter is achieved using a new 3-element corrector optimized in the near-infrared (0.9-1.8um) wavelength range. Due to limited space at the prime-focus, we have had to develop a novel fibre positioner called "Echidna" together with two OH-airglow suppressed spectrographs. FMOS consists of three subsystems: the prime focus unit for IR, the fibre positioning system/connector units, and the two spectrographs. After full systems integration, FMOS was installed on the telescope in late 2007. Many aspects of performance were checked through various test and engineering observations. In this paper, we present the optical and mechanical components of FMOS and show the results of our on-sky engineering observations to date.

Motivation & Objective

  • To develop a wide-field, high-multiplexing near-infrared spectrograph for the Subaru Telescope to enable large-scale galaxy surveys.
  • To overcome the challenge of limited space at the prime focus by designing a novel fiber positioner (Echidna) for 400-object multiplexing.
  • To achieve high spectral efficiency and low background levels in the near-infrared by integrating OH-airglow suppression and thermal blocking filters.
  • To validate system performance through engineering observations and on-sky testing of key subsystems.
  • To enable cosmological and galaxy evolution studies by providing deep, wide-field NIR spectroscopy with high multiplex advantage.

Proposed method

  • FMOS uses a 3-element near-infrared corrector (0.9–1.8 μm) to deliver a 30′ diameter field of view at the prime focus.
  • A novel fiber positioner, Echidna, uses piezo-actuated quadrant tubes to position 400 fibers with sub-10 μm accuracy over 15 minutes.
  • The system employs two cooled infrared spectrographs (IRS1 and IRS2) with VPH gratings and thermal-blocking filters to suppress background.
  • Fiber feed links the prime focus to the spectrographs via 70 m long fibers, maintaining high throughput.
  • The Focal Plane Imager (FPI) on an XY gantry enables iterative fiber positioning and sky image acquisition for field acquisition.
  • OH-airglow suppression is achieved using a mask with a blocking factor of ~10, with potential for further improvement.

Experimental results

Research questions

  • RQ1Can a fiber-fed multi-object spectrograph achieve simultaneous observation of 400 objects within a 30′ field of view in the near-infrared?
  • RQ2What is the achievable configuration accuracy and reconfiguration time of the Echidna fiber positioner system?
  • RQ3To what extent can thermal and OH-airglow backgrounds be suppressed in the spectroscopic system?
  • RQ4What are the limiting magnitudes and signal-to-noise ratios achievable in low-resolution mode for deep surveys?
  • RQ5How does the system efficiency compare to theoretical expectations, and what factors limit performance?

Key findings

  • Echidna achieves 98% of fibers within 10 μm of target positions after 7 iterations, with reconfiguration completed in under 15 minutes.
  • Fiber configuration accuracy measured via raster sampling is less than 0.′′15, and auto-guiding accuracy is below 0.′′18.
  • System throughput from prime focus to detector is 10%, consistent with expectations, though real-world efficiency is 7% due to mismatches and losses.
  • Thermal background rejection in IRS1 approaches ideal performance, indicating effective refrigerator and filter operation.
  • Limiting magnitudes of J = 20.1 mag and H = 19.8 AB mag are achieved for S/N = 5 in a 1-hour integration in low-resolution mode.
  • An emission line flux of 1 × 10⁻¹⁶ erg cm⁻² s⁻¹ is detectable with S/N = 5 in both J and H bands after 1 hour of integration.