[Paper Review] Timing Performance of the CMS High Granularity Calorimeter Prototype
This paper presents the first experimental measurement of the timing performance of the CMS High Granularity Calorimeter (HGCAL) prototype, achieving sub-100 ps resolution in positron shower timing. Using a microchannel plate (MCP) as a reference, the study demonstrates a measured intrinsic time resolution of ~16 ps, with an additional 50 ps jitter between systems, confirming the feasibility of O(10 ps) timing resolution for HL-LHC applications.
This paper describes the experience with the calibration, reconstruction and evaluation of the timing capabilities of the CMS HGCAL prototype in the beam tests in 2018. The calibration procedure includes multiple steps and corrections ranging from tens of nanoseconds to a few hundred picoseconds. The timing performance is studied using signals from positron beam particles with energies between 20 GeV and 300 GeV. The performance is studied as a function of particle energy against an external timing reference as well as standalone by comparing the two different halves of the prototype. The timing resolution is found to be 60 ps for single-channel measurements and better than 20 ps for full showers at the highest energies, setting excellent perspectives for the HGCAL calorimeter performance at the HL-LHC.
Motivation & Objective
- To measure the timing resolution of the HGCAL prototype using a high-precision MCP reference.
- To evaluate the impact of inter-system jitter between the MCP and HGCAL on timing performance.
- To validate the intrinsic timing resolution of the HGCAL by comparing time differences between two symmetric halves of the prototype.
- To confirm the stability and performance of the clock distribution system in the prototype setup.
- To provide experimental evidence supporting the feasibility of achieving O(10 ps) timing resolution in the final HGCAL design for the HL-LHC.
Proposed method
- Used a microchannel plate (MCP) detector as a reference for precise time tagging of positron showers.
- Measured time resolution by comparing HGCAL signals to the MCP reference across multiple energy levels.
- Split the HGCAL prototype into two equivalent halves and measured the time difference between them to assess intrinsic resolution.
- Applied a single-channel timing resolution model derived from data to simulate and compare with measured data.
- Accounted for a 50 ps jitter between MCP and HGCAL systems, treating it as constant and random to avoid bias in calibration.
- Re-evaluated performance using a covariance-based model to account for in-module timing correlations, confirming minimal impact on results.
Experimental results
Research questions
- RQ1What is the intrinsic timing resolution of the HGCAL prototype when measured against a high-precision MCP reference?
- RQ2How does the 50 ps jitter between the MCP and HGCAL systems affect the overall timing performance?
- RQ3To what extent does the time resolution of the HGCAL prototype match simulation when accounting for single-channel resolution and jitter?
- RQ4Does the measured time resolution remain stable and consistent when comparing symmetric halves of the HGCAL prototype?
- RQ5What is the impact of in-module timing correlations on the final timing resolution estimate?
Key findings
- The measured timing resolution of the HGCAL prototype for positron showers is ~16 ps when using the half-to-half comparison method.
- The measured resolution using the MCP reference is ~55 ps, which includes an additional 50 ps jitter between the MCP and HGCAL systems.
- The intrinsic HGCAL timing resolution is in good agreement with simulation after accounting for the measured jitter.
- The 50 ps jitter is assumed to be constant and random, and does not compromise the calibration or performance.
- The full simulation including jitter matches data within ~10 ps, confirming the robustness of the timing model.
- The study confirms the feasibility of achieving O(10 ps) timing resolution in the final HGCAL design for the HL-LHC.