[Paper Review] 3D Augmented Reality-Assisted CT-Guided Interventions: System Design and Preclinical Trial on an Abdominal Phantom using HoloLens 2
This study presents a 3D augmented reality (AR)-assisted navigation system using Microsoft HoloLens 2 to improve CT-guided interventions, particularly for out-of-plane lesions. In a preclinical trial on an abdominal phantom, AR guidance reduced needle passes by 54.2% (from 7.4 to 3.4), lowered radiation dose by 41.0%, and eliminated complications from non-targeted lesions, while standardizing performance across operators of all experience levels.
Background: Out-of-plane lesions pose challenges for CT-guided interventions. Augmented reality (AR) headset devices have evolved and are readily capable to provide virtual 3D guidance to improve CT-guided targeting. Purpose: To describe the design of a three-dimensional (3D) AR-assisted navigation system using HoloLens 2 and evaluate its performance through CT-guided simulations. Materials and Methods: A prospective trial was performed assessing CT-guided needle targeting on an abdominal phantom with and without AR guidance. A total of 8 operators with varying clinical experience were enrolled and performed a total of 86 needle passes. Procedure efficiency, radiation dose, and complication rates were compared with and without AR guidance. Vector analysis of the first needle pass was also performed. Results: Average total number of needle passes to reach the target reduced from 7.4 passes without AR to 3.4 passes with AR (54.2% decrease, p=0.011). Average dose-length product (DLP) decreased from 538 mGy-cm without AR to 318 mGy-cm with AR (41.0% decrease, p=0.009). Complication rate of hitting a non-targeted lesion decreased from 11.9% without AR (7/59 needle passes) to 0% with AR (0/27 needle passes). First needle passes were more nearly aligned with the ideal target trajectory with AR versus without AR (4.6° vs 8.0° offset, respectively, p=0.018). Medical students, residents, and attendings all performed at the same level with AR guidance. Conclusions: 3D AR guidance can provide significant improvements in procedural efficiency and radiation dose savings for targeting challenging, out-of-plane lesions. AR guidance elevated the performance of all operators to the same level irrespective of prior clinical experience.
Motivation & Objective
- To address the clinical challenge of targeting out-of-plane lesions in CT-guided interventions.
- To design and evaluate a 3D augmented reality navigation system using HoloLens 2 for real-time procedural guidance.
- To assess improvements in procedural efficiency, radiation dose, and accuracy across operators of varying experience levels.
- To determine whether AR guidance can standardize performance across medical trainees and attendings.
Proposed method
- Developed a 3D AR navigation system integrating CT imaging data with real-time spatial tracking via HoloLens 2.
- Mapped virtual 3D needle guidance cues onto the physical phantom using fiducial-based registration to align CT and AR views.
- Enabled real-time visualization of the needle path, target, and trajectory in the operator’s field of view through the HoloLens 2 headset.
- Conducted a prospective, blinded trial comparing AR-assisted and conventional CT-guided needle targeting on an abdominal phantom.
- Used vector analysis to quantify alignment of the first needle pass with the ideal trajectory.
- Measured procedural efficiency (number of passes), radiation dose (DLP), and complication rates (non-targeted lesion hits).
Experimental results
Research questions
- RQ1Can 3D AR guidance reduce the number of needle passes required to reach a target in CT-guided interventions?
- RQ2Does AR guidance significantly reduce radiation dose exposure during CT-guided procedures?
- RQ3Can AR guidance improve targeting accuracy, especially for out-of-plane lesions?
- RQ4Does AR guidance reduce the risk of complications such as hitting non-targeted structures?
- RQ5Does AR guidance equalize performance across operators of varying clinical experience levels?
Key findings
- The average number of needle passes decreased from 7.4 without AR to 3.4 with AR, representing a 54.2% reduction (p = 0.011).
- The average dose-length product (DLP) decreased from 538 mGy-cm without AR to 318 mGy-cm with AR, a 41.0% reduction (p = 0.009).
- The complication rate for hitting non-targeted lesions dropped from 11.9% (7 out of 59 passes) without AR to 0% with AR (0 out of 27 passes).
- First needle passes were significantly more aligned with the ideal trajectory under AR guidance (4.6° offset vs. 8.0° without AR, p = 0.018).
- Medical students, residents, and attending physicians all achieved comparable performance levels with AR guidance, indicating performance standardization.
- The system demonstrated consistent improvements across all metrics, regardless of operator experience level.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.