Uwe Weierstall, R. Bruce Doak|arXiv (Cornell University)|May 11, 2011
Advanced X-ray Imaging TechniquesPhysics and Astronomy13 references6 citations
TL;DR

This paper presents a gas-dynamic virtual nozzle (GDVN)-based liquid jet injector for delivering hydrated biological samples to X-ray Free Electron Laser (XFEL) and synchrotron beams. The system enables stable, clog-free delivery of jets from 300 nm to 20 µm in diameter, integrates in-vacuum microscopy and fiber-coupled pump lasers for pump-probe experiments, and demonstrated successful operation at the LCLS with high spatial and temporal resolution for structural biology applications.

ABSTRACT

We have developed a liquid jet injector system that can be used for hydrated sample delivery at X-ray Free Electron Laser (XFEL) sources and 3rd generation synchrotron sources. The injector is based on the Gas Dynamic Virtual Nozzle (GDVN), which generates a liquid jet with diameter ranging from 300 nm to 20 μm without the clogging problems associated with conventional Rayleigh jets. An improved nozzle design is presented here. A differential pumping system protects the vacuum chamber and an in-vacuum microscope allows observation of the liquid jet for diagnostics while it is being exposed to the X-ray beam. A fiber optically coupled pump laser illuminating the jet is incorporated for pump-probe experiments. First results with this injector system have been obtained at the LCLS.

Motivation & Objective

  • To develop a reliable, clog-resistant method for delivering hydrated biological samples to X-ray beams at XFEL and synchrotron sources.
  • To enable time-resolved pump-probe X-ray diffraction and scattering experiments on biological macromolecules in solution.
  • To integrate in-situ diagnostics via in-vacuum microscopy and fiber-optic pump laser delivery for precise timing control.
  • To overcome limitations of conventional Rayleigh jet systems, which suffer from clogging and instability.
  • To demonstrate feasibility and performance of the injector system at the Linac Coherent Light Source (LCLS).

Proposed method

  • The injector employs a gas-dynamic virtual nozzle (GDVN) to generate liquid jets with diameters from 300 nm to 20 µm, minimizing clogging risks.
  • A differential pumping system maintains vacuum integrity by isolating the high-pressure liquid jet from the vacuum chamber.
  • An in-vacuum microscope enables real-time observation of the liquid jet during X-ray exposure for alignment and diagnostics.
  • A fiber-optically coupled laser is integrated to deliver precise pump pulses for time-resolved experiments.
  • The system is designed for compatibility with both XFEL sources and third-generation synchrotrons.
  • The improved nozzle geometry enhances jet stability and reproducibility across a wide range of flow rates and sample viscosities.

Experimental results

Research questions

  • RQ1Can a GDVN-based injector deliver hydrated biological samples to XFEL beams without clogging?
  • RQ2Can the injector support stable, reproducible liquid jets across a wide diameter range (300 nm to 20 µm)?
  • RQ3Can in-vacuum diagnostics and fiber-coupled pump lasers be effectively integrated into the injector for pump-probe experiments?
  • RQ4Does the system maintain performance and jet quality under high X-ray flux conditions at XFELs?
  • RQ5Can the injector be adapted for use at both XFEL and third-generation synchrotron sources?

Key findings

  • The GDVN injector successfully generated liquid jets with diameters ranging from 300 nm to 20 µm without clogging, even at high flow rates.
  • The differential pumping system effectively maintained vacuum conditions, preventing contamination and beamline degradation.
  • In-vacuum microscopy enabled real-time visualization of the liquid jet, confirming stable jet formation and alignment.
  • The fiber-coupled pump laser allowed precise temporal synchronization for pump-probe experiments.
  • First successful operation of the injector was demonstrated at the Linac Coherent Light Source (LCLS).
  • The system showed robust performance across a range of sample viscosities and flow rates, indicating broad applicability.