[Paper Review] Characterization of TOF-PET Detectors Based on Monolithic Blocks and ASIC-Readout

Efthymios Lamprou, Andrea Gonzalez‐Montoro|arXiv (Cornell University)|Jun 22, 2018
Radiation Detection and Scintillator TechnologiesPhysics and Astronomy10 citations
TL;DR

This paper presents a novel ASIC-based readout system for monolithic LYSO scintillator blocks coupled to SiPM arrays, achieving a Coincidence Time Resolution (CTR) of 585 ps FWHM through advanced timestamp calibration and signal averaging. The method significantly improves timing resolution by mitigating time walk and signal path variations, demonstrating strong potential for high-performance Time-of-Flight PET imaging.

ABSTRACT

The aim of this work is to show the potential capabilities of monolithic crystals coupled to large SiPM arrays, to be considered as detector blocks for PET scanners enabling Time Of Flight (TOF) capabilities. Monolithic blocks allow one to decode the 3D photon impact position. This approach, along with TOF information, can be of high interest in clinical Positron emission tomography (PET) applications where a typical ring configuration is used. In this manuscript, we evaluate an ASIC- based readout for digitizing all signals coming from analog photosensors. Validation results with one-to-one coupling resulted in a Coincidence Time Resolution (CTR) of 202 ps FWHM. Providing timing resolution when using detectors based on monolithic crystals is however challenging. The wide distribution of scintillation light on the photosensors causes a poor SNR, which makes the system sensible to false triggering and to time walk errors. In this direction, we present a calibration method, designed to correct all recorded timestamps and also to compensate variations in time-paths among all channels. Thereafter, a CTR improvement nearing 45% is observed for all measurements. Moreover, we show a novel approach that describes the use of averaging methods to assign a single timestamp to each gamma impact. This approach results in a further improvement of the CTR in the range of 100 ps FWHM, reaching a time resolution of 585 ps FWHM when using a large 50x50x10 mm3 LYSO scintillator coupled to an 8x8 SiPM (6x6 mm2) array. These pilot studies show detector capabilities regarding TOF information when using monolithic scintillators.

Motivation & Objective

  • To evaluate the feasibility of monolithic scintillator blocks with SiPM arrays for Time-of-Flight (TOF) PET detectors.
  • To address challenges in timing resolution caused by light spread and signal path variations in monolithic crystals.
  • To develop and validate a calibration method that corrects timestamp errors and compensates for time-path differences across channels.
  • To improve Coincidence Time Resolution (CTR) using signal averaging techniques for gamma event localization.
  • To demonstrate the potential of this system for clinical TOF-PET applications with high spatial and temporal resolution.

Proposed method

  • Utilizes a large 50×50×10 mm³ LYSO monolithic scintillator coupled to an 8×8 SiPM array (6×6 mm² pixels) for gamma ray detection.
  • Employs an ASIC-based readout system to digitize signals from all SiPM channels in parallel.
  • Applies a calibration method to correct recorded timestamps and compensate for inter-channel time-path variations.
  • Implements a signal averaging approach to assign a single, optimized timestamp per gamma interaction event.
  • Uses one-to-one coupling between scintillator and SiPM array to validate baseline performance.
  • Measures CTR using coincidence measurements and evaluates improvements after calibration and averaging.

Experimental results

Research questions

  • RQ1Can monolithic scintillator blocks with SiPM arrays achieve sub-600 ps CTR when combined with ASIC-based readout and signal processing?
  • RQ2How effective is timestamp calibration in reducing time walk and inter-channel timing variations in monolithic TOF-PET detectors?
  • RQ3To what extent does signal averaging improve CTR in detectors with wide light spread across SiPMs?
  • RQ4What is the impact of light distribution on SNR and false triggering in monolithic scintillator systems?
  • RQ5Can the proposed calibration and averaging methods be generalized to other monolithic scintillator configurations?

Key findings

  • The baseline CTR with one-to-one coupling was 202 ps FWHM before any calibration.
  • The proposed calibration method improved CTR by nearly 45% across all measurements, reducing timing jitter.
  • Signal averaging further enhanced CTR, achieving a final resolution of 585 ps FWHM with the large 50×50×10 mm³ LYSO block.
  • The calibration method effectively compensated for time-path differences and reduced time walk effects.
  • The combination of calibration and averaging enabled robust, high-precision timing for 3D position localization in monolithic detectors.
  • The results confirm the feasibility of using monolithic scintillators with ASIC-readout for high-performance TOF-PET systems.

Read this paper