[Paper Review] Experimental validation of Fastcat kV and MV cone beam CT (CBCT) simulator

J. William O’Connell, Clayton Lindsay|arXiv (Cornell University)|Apr 28, 2021
Advanced Radiotherapy TechniquesPhysics and Astronomy34 references8 citations
TL;DR

This paper validates Fastcat, an open-source, fast hybrid cone beam CT (CBCT) simulator, against experimental kV and MV CBCT images from a Varian Truebeam linac. Using detailed models of x-ray beams, detectors, and a high-resolution virtual Catphan phantom, Fastcat achieved sub-15 HU contrast agreement, <4.5% MTF error, <3% CNR RMSE, and <2.5% dose error compared to Monte Carlo and experimental data, with simulations completing in ~70 seconds per scan on a single GPU.

ABSTRACT

Purpose: To experimentally validate the Fastcat cone beam CT (CBCT) simulator against kV and MV CBCT images acquired with a Varian Truebeam linac. Methods: kV and MV CBCT images of a Catphan 504 phantom were acquired using a 100 kVp beam with the on-board imager (OBI) and a 6 MV treatment beam with the electronic portal imaging device (EPID), respectively. The kV Fastcat simulation was performed using detailed models of the x-ray source, bowtie filter, a high resolution voxelized virtual Catphan phantom, anti-scatter grid, and the CsI scintillating detector. Likewise, an MV Fastcat CBCT was simulated with detailed models for the beam energy spectrum, flattening filter, a high resolution voxelized virtual Catphan phantom, and the GOS scintillating detector. Experimental and simulated CBCT images of the phantom were compared with respect to HU values, contrast to noise ratio (CNR),and dose linearity. Detector modulation transfer function (MTF) for the two detectors were also experimentally validated. Fastcat's dose calculations were compared to MC dose calculations performed with Topas. Results: For the kV and MV simulations, respectively: Contrast agreed within 14 and 9 HUs and detector MTF agreed within 4.2% and 2.5%. Likewise, CNR had a root mean squared error (RMSE) of 2.6% and 1.4%. Dose agreed within 2.4% and 1.6% of MC values. The kV and MV CBCT images took 71 and 72 seconds to simulate in Fastcat with 887 and 493 projections, respectively. Conclusions: We present a multi energy experimental validation of a fast and accurate CBCT simulator against a commercial linac. The simulator is open source and all models found in this work can be downloaded from https://github.com/jerichooconnell/fastcat.git

Motivation & Objective

  • To develop and validate a fast, accurate, and open-source CBCT simulator for both kV and MV imaging in radiotherapy.
  • To bridge the gap between clinical patient data (unknown anatomy) and phantom data (limited realism) by enabling simulation on realistic, high-resolution anthropomorphic phantoms with known composition.
  • To achieve simulation speed comparable to analytical methods while maintaining accuracy close to Monte Carlo simulations, particularly for image quality metrics like HU, CNR, and MTF.
  • To enable rapid, reproducible evaluation of new CBCT imaging protocols without patient dose or uncertainty in ground truth.

Proposed method

  • Fastcat uses a hybrid simulation approach combining analytical modeling of beam and detector physics with pre-computed data for efficiency.
  • Detailed models of the x-ray source, bowtie filter (kV), flattening filter (MV), anti-scatter grid, and detector materials (CsI and GOS scintillators) were implemented.
  • A high-resolution voxelized virtual Catphan 504 phantom was used as the anatomical model, with material compositions mapped to Hounsfield Units (HU).
  • Detector response was modeled using experimentally measured optical spread functions (OSFs), accounting for inter-crystal crosstalk beyond nearest neighbors.
  • Simulations were performed using a GPU-accelerated framework, enabling fast projection of 493–887 projections per scan in ~70 seconds.
  • Dose calculations were validated against Monte Carlo (MC) simulations using Topas, with agreement within 2.4% (kV) and 1.6% (MV).

Experimental results

Research questions

  • RQ1Can a fast, open-source CBCT simulator achieve image quality metrics (HU, CNR, MTF) that closely match experimental kV and MV CBCT scans from a commercial linac?
  • RQ2How does the accuracy of Fastcat compare to Monte Carlo simulations in terms of HU, CNR, MTF, and dose for both kV and MV beams?
  • RQ3To what extent does the simulator’s computational efficiency allow for rapid, large-scale virtual CBCT dataset generation suitable for clinical protocol optimization?
  • RQ4Can the simulator accurately model detector spatial resolution (MTF) by incorporating detailed optical spread functions (OSFs) rather than simplified crosstalk models?

Key findings

  • Fastcat achieved HU agreement within 14 HU (kV) and 9 HU (MV) of experimental values, demonstrating high accuracy in Hounsfield Unit representation.
  • Detector MTF values were within 4.2% (kV, CsI) and 2.5% (MV, GOS) of experimental measurements, validating accurate modeling of spatial resolution.
  • Contrast-to-noise ratio (CNR) showed a root mean squared error (RMSE) of 2.6% (kV) and 1.4% (MV) compared to experimental data, indicating strong agreement in image quality.
  • Dose calculations in Fastcat agreed with Monte Carlo (Topas) results within 2.4% (kV) and 1.6% (MV), confirming accurate dose estimation.
  • Simulations completed in 71 seconds (kV) and 72 seconds (MV) on a single GPU, with 887 and 493 projections respectively, demonstrating high computational efficiency.
  • The simulator is open-source and publicly available at https://github.com/jerichooconnell/fastCAT.git, with all models and pre-computed data for the Varian Truebeam system included.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.