[Paper Review] Improvement of radon detector performance by using a large-sized PIN-photodiode
This paper proposes enhancing radon detector sensitivity by replacing a small (18×18 mm) PIN-photodiode with a larger (28×28 mm) one, improving detection efficiency by 3.8±2.4% at low radon concentrations (<1.0 g/m³) and up to 10–20% at higher levels, without increasing intrinsic background. The upgrade enables more precise radon measurements in humid gases and material screening, crucial for rare-event experiments.
Radioactive noble gas radon ($\mathrm{^{222}Rn}$) is one of the major background sources below the MeV region in rare event search experiments. To precisely measure radon concentration in purified gases, a radon detector with an electrostatic collection method is widely used. In this paper, we discussed the improvements of a radon detector by installing a new PIN-photodiode ($28 imes28$ $\mathrm{mm}$) whose surface area is $2.5$ times larger than that used previously ($18 imes18$ $\mathrm{mm}$). We evaluated the detector's performance by serially connecting two radon detectors equipped with two types of PIN-photodiodes. As a result of the calibrations, we found an improvement of $(3.8\pm2.4)\%$ in the detection efficiencies below $1.0$ $\mathrm{g/m^{3}}$, while a $10$-$20\%$ improvement occurred above this level. The intrinsic background of the detector equipped with the large PIN-photodiode was measured as $0.24^{+0.09}_{-0.05}$ $\mathrm{mBq/m^{3}}$. This background level is consistent with the radon detector with the small PIN-photodiode, although we installed the large one. This improvement is useful for applications in radon emanation measurements from a material, which also emits water from its surface.
Motivation & Objective
- To improve the detection sensitivity of radon detectors used in rare-event experiments, particularly for measuring low-level radon in purified gases.
- To address the challenge of reduced detection efficiency in humid environments, where water vapor neutralizes radon daughter ions.
- To evaluate the impact of increasing the PIN-photodiode surface area on detection efficiency and intrinsic background.
- To enable more accurate radon emanation rate measurements from materials used in detector construction.
- To support real-time monitoring of radon in purified water and gases, essential for background suppression in large-scale experiments.
Proposed method
- Replaced the standard 18×18 mm PIN-photodiode with a larger 28×28 mm photodiode in a radon detector with an electrostatic collection system.
- Constructed a calibration setup by serially connecting two identical radon detectors, each equipped with one of the two photodiode types, to compare performance.
- Measured the high-voltage dependence of the calibration factor to assess collection efficiency under varying electric fields.
- Evaluated the absolute humidity dependence of the calibration factor to test performance in humid gas conditions.
- Calculated the detection efficiency correction factor ε = 0.1659 based on the 3.82-day half-life of ²²²Rn to account for decay during emanation.
- Measured the intrinsic background of the large-diode detector as 0.24⁺⁰·⁰⁹₋₀·⁰⁵ mBq/m³ to assess background stability.
Experimental results
Research questions
- RQ1Does increasing the PIN-photodiode surface area from 18×18 mm to 28×28 mm improve the detection efficiency of a radon detector in low-concentration and humid gas environments?
- RQ2What is the impact of the larger photodiode on the intrinsic background level of the detector?
- RQ3How does the calibration factor of the detector vary with applied high voltage and absolute humidity?
- RQ4To what extent does the larger photodiode enhance the detector's sensitivity for measuring radon emanation from materials?
- RQ5Can the upgraded detector maintain low background while achieving higher efficiency in real-time radon monitoring applications?
Key findings
- The detection efficiency improved by (3.8±2.4)% at radon concentrations below 1.0 g/m³ due to the larger photodiode.
- At higher concentrations (>1.0 g/m³), the improvement in detection efficiency reached 10–20%.
- The intrinsic background of the detector with the large PIN-photodiode was measured as 0.24⁺⁰·⁰⁹₋₀·⁰⁵ mBq/m³, consistent with the previous small-diode design.
- The calibration factor showed a 15–30% improvement at high voltages above -0.3 kV, indicating better charge collection efficiency.
- The upgrade demonstrated enhanced performance in humid gas conditions, crucial for applications like radon monitoring in purified water.
- The study confirms that increasing photodiode size is an effective, low-cost method to improve detector sensitivity without increasing background.
Read this paper