[Paper Review] IoT-Based Secure Embedded Scheme for Insulin Pump Data Acquisition and Monitoring
This paper proposes an IoT-based embedded system using a Keil LPC1768 microcontroller to securely acquire and monitor insulin pump data, applying SHA-256 hashing to ensure data integrity and authenticity. The system verifies data consistency between cloud and patient-side records, detecting any tampering through hash comparison, and ensures secure command transmission to the infusion pump.
This paper introduces an Internet of Things (IoT)-based data acquisition and monitoring scheme for insulin pumps. The proposed work employs embedded system hardware (Keil LPC1768-board) for data acquisition and monitoring. The hardware is used as an abstract layer between the insulin pump and the cloud. Diabetes data are secured before they are sent to the cloud for storage. Each patient's record is digitally signed using a secure hash algorithm mechanism. The proposed work will protect the patient's records from being breached from unauthorized entities, and authenticates them from improper modifications. The design is tested and verified using $μ$Vision studio, the Keil board mentioned above, and an ALARIS 8100 infusion pump. Moreover, a test case for a real cloud example is presented with the help of the Center of Computationally Assisted System and Technology. This center provided the infrastructure service to test our work.
Motivation & Objective
- To address security vulnerabilities in insulin pump data transmission and storage by implementing a lightweight, hardware-based authentication mechanism.
- To ensure data integrity and prevent unauthorized modification of patient health records during cloud-based transmission.
- To enable real-time, secure monitoring of diabetes-related data using an embedded system as an abstraction layer between the insulin pump and cloud.
- To validate the system’s ability to detect data tampering through cryptographic hash verification in a real-world test environment.
- To support future expansion of patient control over insulin delivery schedules while maintaining end-to-end security.
Proposed method
- The Keil LPC1768 microcontroller acts as an embedded hardware abstraction layer between the Alaris 8100 insulin pump and the cloud.
- Patient health data, including glucose levels and medical profiles, are collected and secured using the SHA-256 cryptographic hash function before transmission.
- Each data record is digitally signed with a SHA-256 hash value, which is appended to the data packet for integrity verification.
- Upon receiving data, the system recomputes the SHA-256 hash on the patient side and compares it with the transmitted hash; mismatched values indicate tampering.
- Control commands sent from the hospital to the pump are similarly signed with SHA-256 and verified by the microcontroller before execution.
- The system leverages fault exception handling in the Cortex-M3 processor to detect and respond to hardware-level errors during data processing.
Experimental results
Research questions
- RQ1How can an embedded system securely mediate data exchange between an insulin pump and the cloud while preserving data integrity?
- RQ2To what extent can SHA-256 hashing detect unauthorized modifications to patient health records during transmission?
- RQ3Can a low-power microcontroller like the Keil LPC1768 effectively enforce cryptographic authentication in a real-time medical IoT environment?
- RQ4How does the system ensure that only authenticated and unaltered control commands are executed by the insulin pump?
- RQ5What is the effectiveness of the proposed scheme in detecting data tampering in a real-world test setup with actual medical devices?
Key findings
- The system successfully detected data tampering by comparing SHA-256 hash values between the cloud and patient-side systems, with mismatched hashes indicating data alteration.
- In Table I, identical SHA-256 hash values (14b93acf-ccdcbe40-ea3795be-c1073498-51a96c90-6cedfc9c-49d8e2cf-a141befb) on both cloud and patient sides confirmed data integrity.
- In Table II, a mismatched hash value (358c4f29-f0e2bb60-8efa35d4-a88a6b3b-58939ffd-deebf824-8065c195-b834b8cd) on the patient side indicated data modification, which the system correctly flagged and discarded.
- The Keil LPC1768 microcontroller successfully verified the authenticity of control commands by recomputing and comparing SHA-256 values, ensuring only valid commands were passed to the infusion pump.
- The system demonstrated real-time integrity verification using actual data from a 24-hour glucose monitoring period, confirming consistent and secure data handling.
- The integration with the Center of Computationally Assisted System and Technology provided a validated cloud infrastructure for testing, confirming the scheme’s feasibility in a production-like environment.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.