[Paper Review] Proton radiographs using position-sensitive silicon detectors and high-resolution scintillators

J. A. Briz, A. N. Nerio|arXiv (Cornell University)|Nov 16, 2021
Radiation Therapy and DosimetryMedicine16 references4 citations
TL;DR

This study presents a prototype proton radiography system using position-sensitive silicon detectors (DSSDs) and a high-resolution scintillator array to enable accurate proton imaging for cancer therapy. The system achieved sub-2 mm spatial resolution and distinguished 10-mm thicknesses of aluminum and PMMA, validating its potential as a first step toward proton CT for improved treatment planning and dose verification.

ABSTRACT

Proton therapy is a cancer treatment technique currently in growth worldwide. It offers advantages with respect to conventional X-ray and $\gamma$-ray radiotherapy, in particular, a better control of the dose deposition allowing to reach a higher conformity in the treatments. Therefore, it causes less damage to the surrounding healthy tissue and less secondary effects. However, in order to take full advantage of its potential, improvements in treatment planning and dose verification are required. A new prototype of proton Computed Tomography scanner is proposed to design more accurate and precise treatment plans for proton therapy. Here, results obtained from an experiment performed using a 100-MeV proton beam at the CCB facility in Krakow (Poland) are presented. Proton radiographs of PMMA samples of 50-mm thickness with spatial patterns in aluminum were taken. Their properties were studied, including reproduction of the dimensions, spatial resolution and sensitivity to different materials. They demonstrate the capabilities of the system to produce images with protons. Structures of up to 2 mm are nicely resolved and the sensitivity of the system was enough to distinguish thicknesses of 10 mm of aluminum or PMMA. This constitutes a first step to validate the device as a proton radiography scanner previous to the future tests as a proton CT scanner.

Motivation & Objective

  • To develop and validate a proton radiography system using position-sensitive silicon detectors and scintillators for improved proton therapy imaging.
  • To address uncertainties in proton range estimation caused by converting Hounsfield Units from X-ray CT to Relative Stopping Powers (RSP).
  • To demonstrate the feasibility of using protons for imaging with high spatial and energy resolution, reducing reliance on X-ray CT-based planning.
  • To validate the system's ability to resolve fine structures and distinguish material thicknesses relevant to clinical proton therapy.
  • To lay the groundwork for future proton CT reconstruction by testing image quality and detector performance under clinical-relevant beam conditions.

Proposed method

  • Used a 90–120 MeV proton beam at the CCB facility in Krakow, Poland, with beam energy degraded from 230 MeV.
  • Employed two double-sided silicon strip detectors (DSSDs) for precise proton tracking and trajectory reconstruction.
  • Integrated a lanthanum bromide (LaBr3) scintillator array to measure residual proton energy and improve stopping power determination.
  • Performed energy calibration using 95, 100, and 120 MeV proton beams and validated with detailed Monte Carlo simulations.
  • Acquired proton radiographs of PMMA samples with embedded aluminum patterns of varying sizes (1–5 mm) and spacings.
  • Analyzed images using 1D intensity profiles and super-Gaussian fits to assess spatial resolution and dimensional accuracy.

Experimental results

Research questions

  • RQ1Can a proton radiography system using DSSDs and scintillators resolve structures down to 2 mm in size?
  • RQ2What is the system's sensitivity to thickness differences in materials like aluminum and PMMA?
  • RQ3How accurately can the system reproduce the true dimensions of embedded test patterns in PMMA?
  • RQ4To what extent does the spatial resolution degrade for smaller features (e.g., 1 mm holes)?
  • RQ5Can the system distinguish 10-mm thicknesses of different materials based on residual energy deposition?

Key findings

  • The system achieved a spatial resolution better than 2 mm, successfully resolving 5-mm, 3-mm, and 2-mm diameter holes in aluminum patterns.
  • The 1-mm holes were not clearly resolved, indicating the system’s resolution limit is between 1 mm and 2 mm.
  • Three distinct energy peaks were observed in the residual energy spectrum, confirming the system’s ability to distinguish 10-mm thicknesses of aluminum, PMMA, and air.
  • The measured dimensions of the cross-shaped and regular pattern samples were in good agreement with the actual physical dimensions, with fits to super-Gaussian functions showing high fidelity.
  • The intensity profiles of the 5-mm, 3-mm, and 2-mm hole rows showed clear minima corresponding to the number of holes, confirming structural resolution.
  • The results validate the prototype as a functional proton radiography scanner and a critical first step toward full proton CT implementation.