[Paper Review] STTAR: Surgical Tool Tracking using off-the-shelf Augmented Reality Head-Mounted Displays
This paper presents STTAR, a framework that leverages the built-in Time-of-Flight (ToF) camera in off-the-shelf AR headsets like Microsoft HoloLens 2 to track retro-reflective surgical tool markers with sub-millimeter accuracy, enabling real-time, multi-tool tracking without external hardware. The method achieves 0.09±0.06 mm lateral translation, 0.42±0.32 mm longitudinal translation, and 0.80±0.39° rotation accuracy, demonstrating feasibility for AR-guided k-wire insertion in orthopedic procedures.
The use of Augmented Reality (AR) for navigation purposes has shown beneficial in assisting physicians during the performance of surgical procedures. These applications commonly require knowing the pose of surgical tools and patients to provide visual information that surgeons can use during the task performance. Existing medical-grade tracking systems use infrared cameras placed inside the Operating Room (OR) to identify retro-reflective markers attached to objects of interest and compute their pose. Some commercially available AR Head-Mounted Displays (HMDs) use similar cameras for self-localization, hand tracking, and estimating the objects' depth. This work presents a framework that uses the built-in cameras of AR HMDs to enable accurate tracking of retro-reflective markers, such as those used in surgical procedures, without the need to integrate any additional components. This framework is also capable of simultaneously tracking multiple tools. Our results show that the tracking and detection of the markers can be achieved with an accuracy of 0.09 +- 0.06 mm on lateral translation, 0.42 +- 0.32 mm on longitudinal translation, and 0.80 +- 0.39 deg for rotations around the vertical axis. Furthermore, to showcase the relevance of the proposed framework, we evaluate the system's performance in the context of surgical procedures. This use case was designed to replicate the scenarios of k-wire insertions in orthopedic procedures. For evaluation, two surgeons and one biomedical researcher were provided with visual navigation, each performing 21 injections. Results from this use case provide comparable accuracy to those reported in the literature for AR-based navigation procedures.
Motivation & Objective
- To develop a method for accurate surgical tool tracking using only the built-in cameras of commercial AR head-mounted displays (HMDs), eliminating the need for additional tracking hardware.
- To address the challenge of maintaining line-of-sight between surgical markers and tracking systems in cluttered operating rooms by leveraging HMD-based tracking.
- To evaluate the feasibility of using AR HMDs for real-time, multi-tool tracking in a clinical-relevant scenario, such as k-wire insertion in orthopedic surgery.
- To compare the performance of the proposed method with existing marker-based tracking systems and assess its potential for novice users.
Proposed method
- Utilizes the built-in Time-of-Flight (ToF) sensor of the Microsoft HoloLens 2 to detect and localize passive retro-reflective markers attached to surgical tools.
- Employs a calibration model that aligns the HMD’s view with the image sensors to ensure accurate spatial registration between virtual and real-world coordinates.
- Applies computer vision algorithms to detect and track retro-reflective markers in real time, leveraging the depth and spatial resolution of the ToF sensor.
- Integrates tool tracking data with the HMD’s self-localization data to compute the 3D pose of surgical tools relative to the patient and environment.
- Uses a virtual trajectory overlay in the HMD to guide surgeons during k-wire insertion, based on the tracked tool pose.
- Implements a multi-target tracking pipeline that maintains a frame rate above 50 Hz even when tracking five different tools with multiple markers simultaneously.
Experimental results
Research questions
- RQ1Can the built-in ToF camera of an off-the-shelf AR HMD achieve sufficient accuracy and stability for surgical tool tracking in a clinical environment?
- RQ2How does the tracking accuracy of the proposed method compare to that of commercial medical-grade tracking systems in terms of translation and rotation?
- RQ3Can the framework support real-time, multi-tool tracking without additional hardware or external illumination?
- RQ4What is the performance of the system in a realistic surgical task, such as k-wire insertion, and how does it compare to existing AR navigation systems?
- RQ5Does the system enable comparable accuracy for novice users compared to experienced surgeons in AR-guided procedures?
Key findings
- The proposed framework achieves a tracking accuracy of 0.09±0.06 mm for lateral translation, 0.42±0.32 mm for longitudinal translation, and 0.80±0.39° for rotation around the vertical axis.
- The system maintains a tracking frame rate above 172 Hz for a single tool and remains above 50 Hz when tracking five tools with multiple markers simultaneously.
- The method outperforms traditional feature-based tracking techniques such as ArUco (35 Hz) and ChArUco (33.2 Hz) in frame rate, and exceeds the AHAT camera’s frame rate (45 Hz).
- The system demonstrated comparable accuracy to literature-reported AR-based navigation systems in a k-wire insertion task, with no significant difference in performance between experienced surgeons and a biomedical researcher.
- The latency in virtual content rendering remains a key challenge, particularly over wireless networks with frame rates below 12 fps, due to potential desynchronization between tracking and HMD localization data.
- The method does not require external infrared illumination, avoiding interference with the HMD’s hand tracking and environmental reconstruction functionalities.