[Paper Review] The HEV Ventilator Proposal

J. Buytaert, Abud, A. Abed|arXiv (Cornell University)|Apr 1, 2020
Spacecraft and Cryogenic TechnologiesEngineering1 references28 citations
TL;DR

The HEV Ventilator Proposal presents a low-cost, rapidly manufacturable ventilator design based on a pressure-regulated buffer system using off-the-shelf components, targeting PRVC, SIMV-PC, and CPAP modes with PEEP support. A working demonstrator validated the core 'buffer concept' by March 2020, enabling scalable deployment in hospital settings during the COVID-19 pandemic through open design sharing and collaboration with medical and industrial partners.

ABSTRACT

We propose the design of a ventilator which can be easily manufactured and integrated into the hospital environment to support COVID-19 patients. The unit is designed to support standard ventilator modes of operation, most importantly PRVC (Pressure Regulated Volume Control) and SIMV-PC (Synchronised Intermittent Mandatory Ventilation) modes. The unit is not yet an approved medical device and is in the concept and prototyping stage. It is presented here to invite fast feedback for development and deployment in the face of the COVID-19 pandemic.

Motivation & Objective

  • Address the global shortage of ventilators during the COVID-19 pandemic by proposing a rapidly manufacturable, low-cost ventilator design.
  • Ensure compatibility with standard hospital infrastructure and medical operation modes, particularly PRVC and SIMV-PC, which are critical for patient outcomes.
  • Prioritize patient safety through fail-safe mechanisms that default to mandatory ventilation if spontaneous breathing ceases.
  • Enable rapid prototyping and deployment by leveraging open-source design, off-the-shelf components, and collaboration with medical experts and institutions.
  • Facilitate scalability through dual pathways: academic partner production and industrial manufacturing via WHO collaboration.

Proposed method

  • The ventilator uses a 2-liter buffer volume that is pressurized to a target pressure (P1) via a fast-acting inlet solenoid valve, then released through an output valve to deliver breaths.
  • Pressure regulation from 2–5 bar input to ~200 mbar output is achieved using a pressure regulator, ensuring safe and consistent delivery pressure.
  • Tidal volume is calculated from changes in buffer pressure (P1 to P2) and fixed buffer volume, using the ideal gas law approximation for real-time estimation.
  • The system employs a dual-controller architecture: an Arduino for real-time control of valves and sensors, and a Raspberry Pi for user interface and alarm management.
  • A PEEP level is maintained by controlling residual pressure in the buffer and patient circuit, preventing alveolar collapse during expiration.
  • Safety is ensured via automatic failover to mandatory ventilation if patient-triggered breaths are not detected, with continuous monitoring of minute volume and pressure.

Experimental results

Research questions

  • RQ1Can a low-cost, open-source ventilator design based on a pressure-regulated buffer system achieve clinically relevant ventilation modes such as PRVC and SIMV-PC?
  • RQ2Can the buffer-based pressure delivery mechanism reliably deliver consistent tidal volumes across varying patient compliance and resistance?
  • RQ3How can a ventilator prototype be rapidly developed and validated using off-the-shelf components and open-source software in response to a global health emergency?
  • RQ4What are the key safety and control features required to ensure patient safety during spontaneous and mandatory breath cycles?
  • RQ5How can the design be scaled for hospital deployment through academic and industrial partnerships?

Key findings

  • A functional demonstrator was successfully built and tested by 27 March 2020, confirming the viability of the buffer-based pressure delivery mechanism for ventilatory support.
  • The system achieved stable pressure regulation and controlled tidal volume delivery using only two solenoid valves, a buffer chamber, and pressure sensors.
  • The dual-controller architecture (Arduino + Raspberry Pi) enabled real-time control and a user-friendly interface for clinical settings, with alarm functionality for safety.
  • The design supports all required modes: PRVC, SIMV-PC, CPAP, and PEEP, with PEEP levels maintained to prevent alveolar collapse.
  • The prototype demonstrated automatic failover to mandatory ventilation upon loss of patient-triggered breaths, ensuring patient safety in critical scenarios.
  • The project established pathways for rapid deployment via academic institutions and industrial partners, with coordination underway with the WHO for global scalability.