[Paper Review] Evaluation of Peak-Fitting Software for Gamma Spectrum Analysis

Guilherme S. Zahn, Frederico A. Genezini|arXiv (Cornell University)|Nov 13, 2015
Nuclear Physics and ApplicationsPhysics and Astronomy2 references10 citations
TL;DR

This study evaluates the performance of five automatic gamma spectrum analysis software—Genie 2000, GammaVision, HypermetPC, SAANI, and VISPECT—against a manual peak-fitting tool (IDeFix) using standard 137Cs, 60Co, 133Ba, and 152Eu sources. All automatic software performed reliably except GammaVision, which failed to identify key peaks and underestimated peak areas; HypermetPC and Genie 2000 showed the best overall accuracy, with automatic tools matching or exceeding manual fitting in precision and speed.

ABSTRACT

In all applications of gamma-ray spectroscopy, one of the most important and delicate parts of the data analysis is the fitting of the gamma-ray spectra, where information as the number of counts, the position of the centroid and the width, for instance, are associated with each peak of each spectrum. There's a huge choice of computer programs that perform this type of analysis, and the most commonly used in routine work are the ones that automatically locate and fit the peaks; this fit can be made in several different ways -- the most common ways are to fit a Gaussian function to each peak or simply to integrate the area under the peak, but some software go far beyond and include several small corrections to the simple Gaussian peak function, in order to compensate for secondary effects. In this work several gamma-ray spectroscopy software are compared in the task of finding and fitting the gamma-ray peaks in spectra taken with standard sources of $^{137}$Cs, $^{60}$Co, $^{133}$Ba and $^{152}$Eu. The results show that all of the automatic software can be properly used in the task of finding and fitting peaks, with the exception of GammaVision; also, it was possible to verify that the automatic peak-fitting software did perform as well as -- and sometimes even better than -- a manual peak-fitting software.

Motivation & Objective

  • To evaluate the reliability and accuracy of automatic gamma spectrum analysis software in identifying and fitting gamma-ray peaks.
  • To compare automatic software performance against a manual peak-fitting tool (IDeFix) in terms of peak area determination and uncertainty estimation.
  • To assess software capabilities in resolving overlapping or closely spaced peaks, such as doublets in 152Eu and 133Ba.
  • To determine whether automatic software can achieve results comparable or superior to manual analysis in routine analytical spectrometry.
  • To update and extend the 1997 IAEA intercomparison by testing newer software versions and updated methodologies.

Proposed method

  • Experimental gamma spectra were acquired using high-purity germanium detectors with standard sources (137Cs, 60Co, 133Ba, 152Eu) over 3600 s live time.
  • Peak fitting was performed using five automatic software tools (Genie 2000, GammaVision, HypermetPC, SAANI, VISPECT) with default settings, except for HypermetPC, where the 'Try' parameter was reduced to 2 to reduce false peak detection.
  • The Gaussian function was used as the primary peak shape model: $ Y(x) = A \cdot e^{-(x-\bar{x})^2 / (2s^2)} $, with background modeled by low-order polynomials.
  • Peak area was calculated as net area: $ N = A - B \cdot (W_p / W_b) $, with uncertainty $ \sigma_N = \sqrt{A + B \cdot (W_p / W_b)^2} $, and fit-based uncertainties were also evaluated.
  • Results were validated using Z-scores and percent uncertainties for peak areas, comparing software outputs to certified values from standard sources.
  • The performance was assessed on three key challenges: fitting the 662 keV peak in 137Cs (high intensity, non-Gaussian shape), resolving the 1086–1090 keV doublet in 152Eu, and identifying the 80–81 keV doublet in 133Ba.

Experimental results

Research questions

  • RQ1How accurately do automatic gamma spectrum analysis software determine peak areas compared to certified values for standard sources?
  • RQ2Can automatic software reliably resolve closely spaced or overlapping peaks, such as the 1086–1090 keV doublet in 152Eu?
  • RQ3Why does GammaVision consistently underperform in peak detection and area estimation compared to other software?
  • RQ4To what extent do automatic software tools match or exceed the accuracy of manual peak-fitting methods in terms of area determination and uncertainty estimation?
  • RQ5How do different software tools handle non-ideal peak shapes, such as the distorted 662 keV peak from 137Cs?

Key findings

  • GammaVision failed to identify several key peaks and significantly underestimated peak areas, particularly for the 662 keV peak in 137Cs, with a Z-score of -3.24 and area uncertainty of 0.051%.
  • HypermetPC was the only software to correctly identify both 80 keV and 81 keV peaks in the 133Ba source doublet, though it reported high uncertainties (5.17% for 80 keV).
  • Genie 2000 performed well across all tests, achieving Z-scores within ±1.0 for all major peaks and reporting low uncertainties (e.g., 0.29% for 1086 keV in 152Eu).
  • IDeFix, the manual fitting software, performed comparably to the best automatic tools, but required significantly more time; automatic software achieved similar or better precision in a fraction of the time.
  • The 662 keV peak from 137Cs, known for non-Gaussian shape due to pile-up, was best fitted by HypermetPC and IDeFix, which reported Z-scores of -0.01 and 0.39, respectively, indicating minimal deviation from certified values.
  • SAANI and VISPECT showed mixed performance: SAANI failed to resolve the 80–81 keV doublet, while VISPECT struggled with the 1090 keV peak in 152Eu, yielding Z-scores of -3.76 and 1.00, respectively.