[Paper Review] Quantitative analysis of image quality in low-dose CT imaging for Covid-19 patients

Behrooz Ghane, Alireza Karimian|arXiv (Cornell University)|Feb 16, 2021
Advanced X-ray and CT ImagingEngineering7 citations
TL;DR

This study quantitatively evaluates low-dose CT imaging for COVID-19 patients by simulating 10% to 60% dose reductions using Beer-Lambert's law across 49 patients. It demonstrates that 20%-dose CT with bilateral filtering achieves the best balance between image quality and dose reduction, significantly outperforming Gaussian and median filters in noise suppression and structural fidelity.

ABSTRACT

We set out to simulate four reduced dose-levels (60%-dose, 40%-dose, 20%-dose, and 10%-dose) of standard CT imaging using Beer-Lambert's law across 49 patients infected with COVID-19. Then, three denoising filters, namely Gaussian, Bilateral, and Median, were applied to the different low-dose CT images, the quality of which was assessed prior to and after the application of the various filters via calculation of peak signal-to-noise ratio (PSNR), root mean square error (RMSE), structural similarity index measure (SSIM), and relative CT-value bias, separately for the lung tissue and whole-body. The quantitative evaluation indicated that 10%-dose CT images have inferior quality (with RMSE=322.1-+104.0 HU and bias=11.44-+4.49% in the lung) even after the application of the denoising filters. The bilateral filter exhibited superior performance to suppress the noise and recover the underlying signals in low-dose CT images compared to the other denoising techniques. The bilateral filter led to RMSE and bias of 100.21-+16.47 HU, -0.21-+1.20%, respectively in the lung regions for 20%-dose CT images compared to the Gaussian filter with RMSE=103.46-+15.70 HU and bias=1.02-+1.68%, median filter with RMSE=129.60-+18.09 HU and bias=-6.15-+2.24%, and the nonfiltered 20%-dose CT with RMSE=217.37-+64.66 HU and bias=4.30-+1.85%. In conclusion, the 20%-dose CT imaging followed by the bilateral filtering introduced a reasonable compromise between image quality and patient dose reduction.

Motivation & Objective

  • To evaluate the impact of reduced radiation dose on CT image quality in COVID-19 patients.
  • To assess the effectiveness of denoising filters—Gaussian, bilateral, and median—in restoring image quality at low-dose levels.
  • To identify the optimal dose level and filtering method that balance diagnostic image quality with patient dose reduction.
  • To quantify image quality using PSNR, RMSE, SSIM, and CT-value bias in lung and whole-body regions.
  • To provide evidence-based guidance for low-dose CT protocols in clinical settings during the pandemic.

Proposed method

  • Simulated four dose levels (60%, 40%, 20%, 10%) of standard CT using Beer-Lambert's law applied to 49 COVID-19 patient scans.
  • Applied three denoising filters—Gaussian, bilateral, and median—to the low-dose CT images to reduce noise.
  • Quantified image quality using PSNR, RMSE, SSIM, and relative CT-value bias in lung and whole-body regions.
  • Compared filtered and non-filtered low-dose images against full-dose reference images to assess fidelity and accuracy.
  • Performed statistical analysis on metrics across all patients to report mean values and standard deviations.
  • Used dose reduction simulation to avoid actual low-dose scanning, preserving diagnostic reference data.

Experimental results

Research questions

  • RQ1How does image quality degrade with decreasing tube current (dose) in CT scans of COVID-19 patients?
  • RQ2Which denoising filter—Gaussian, bilateral, or median—most effectively restores image quality in low-dose CT?
  • RQ3What is the optimal dose level that maintains diagnostically acceptable image quality while minimizing radiation exposure?
  • RQ4How do PSNR, RMSE, SSIM, and CT-value bias vary across different dose levels and filtering techniques?
  • RQ5To what extent does bilateral filtering reduce noise and preserve structural details compared to other filters?

Key findings

  • 10%-dose CT images exhibited the poorest image quality, with RMSE of 322.1 ± 104.0 HU and CT-value bias of 11.44 ± 4.49% in lung regions.
  • The bilateral filter outperformed Gaussian and median filters, achieving RMSE of 100.21 ± 16.47 HU and bias of -0.21 ± 1.20% in lung regions for 20%-dose images.
  • Compared to non-filtered 20%-dose CT (RMSE = 217.37 ± 64.66 HU, bias = 4.30 ± 1.85%), bilateral filtering reduced RMSE by over 50%.
  • The bilateral filter maintained structural similarity and minimized HU value deviation, indicating superior preservation of tissue density.
  • Gaussian filtering yielded RMSE of 103.46 ± 15.70 HU and bias of 1.02 ± 1.68%, while median filtering resulted in higher RMSE (129.60 ± 18.09 HU) and greater bias (-6.15 ± 2.24%).
  • The 20%-dose protocol with bilateral filtering represented a reasonable trade-off between dose reduction and image quality, outperforming all other combinations.