[Paper Review] Simple Self-calibrating Polarimeter for Measuring the Stokes Parameters of Light

Vitaly Wirthl, Cristian D. Panda|arXiv (Cornell University)|May 26, 2021
Optical Polarization and EllipsometryEngineering27 references9 citations
TL;DR

This paper presents a simple, self-calibrating rotating-waveplate polarimeter that accurately measures Stokes parameters of light with minimal sensitivity to intensity fluctuations. By using in situ calibration with a linearly polarized source and dual detectors, it achieves 0.1% accuracy in circular polarization and 0.4% in linear polarization fractions, enabling precise polarization gradient measurements critical for the ACME electron electric dipole moment experiment.

ABSTRACT

A simple, self-calibrating, rotating-waveplate polarimeter is largely insensitive to light intensity fluctuations and is shown to be useful for determining the Stokes parameters of light. This study shows how to minimize the in situ self-calibration time, the measurement time and the measurement uncertainty. The suggested methods are applied to measurements of spatial variations in the linear and circular polarizations of laser light passing through glass plates with a laser intensity dependent birefringence. These are crucial measurements for the ACME electron electric dipole measurements, requiring accuracies in circular and linear polarization fraction of about 0.1% and 0.4%, with laser intensities up to 100 $ ext{mW/mm}^2$ incident into the polarimeter.

Motivation & Objective

  • To develop a low-cost, robust polarimeter for measuring Stokes parameters of partially polarized light in experimental physics.
  • To address systematic uncertainties in electron electric dipole moment measurements caused by polarization gradients from birefringent glass plates.
  • To minimize measurement time, uncertainty, and calibration duration in polarimetry without removing or reconfiguring optical elements.
  • To enable accurate in situ calibration using only a linearly polarized calibration source and dual detectors, avoiding external calibration steps.
  • To achieve high-precision polarization measurements under high laser intensities (up to 100 mW/mm²) with minimal sensitivity to intensity fluctuations.

Proposed method

  • The polarimeter uses a rotating quarter-wave plate followed by a rotating linear polarizer and a detector, both independently rotatable about their axes.
  • It employs a dual-detector setup where one detector rotates with the linear polarizer, reducing sensitivity to incident light intensity fluctuations.
  • The system uses in situ calibration with a linearly polarized light source to determine unknown parameters such as waveplate retardance δ, and angular offsets α₀ and β₀.
  • The detected intensity is modeled using Mueller matrices for the waveplate and polarizer, with the full Stokes vector reconstructed from intensity measurements at varying rotation angles.
  • The method accounts for imperfect polarizer extinction ratio r via a modified Mueller matrix, assuming r ≲ 10⁻⁵ for high dynamic range.
  • Calibration is performed entirely in situ without disassembling or reconfiguring the device, ensuring practicality and repeatability.

Experimental results

Research questions

  • RQ1How can a simple polarimeter be designed to minimize sensitivity to light intensity fluctuations during Stokes parameter measurement?
  • RQ2What is the minimal set of in situ measurements required to self-calibrate the angular offsets and waveplate retardance in a rotating-waveplate polarimeter?
  • RQ3How can measurement time and uncertainty be minimized while maintaining sub-0.1% accuracy in circular and 0.4% in linear polarization fractions?
  • RQ4To what extent can a high-extinction-ratio polarizer and in situ calibration reduce systematic errors from birefringence in optical components?
  • RQ5Can this polarimeter achieve the required precision for the ACME experiment under high laser intensities (up to 100 mW/mm²) and complex polarization gradients?

Key findings

  • The polarimeter achieves a circular polarization fraction measurement accuracy of approximately 0.1%.
  • The linear polarization fraction is measured with an uncertainty of about 0.4%.
  • In situ calibration using a linearly polarized source eliminates the need for external calibration or reconfiguration of the device.
  • The dual-detector configuration significantly reduces sensitivity to intensity fluctuations in the incident light.
  • The method enables accurate measurement of spatial polarization gradients in laser beams passing through birefringent glass plates.
  • The system is robust against residual circular polarization contamination from imperfect polarizers, with contamination levels below √r ≈ 10⁻³ for typical high-quality polarizers.