[Paper Review] An InGrid based Low Energy X-ray Detector

C. Krieger, Klaus Desch|arXiv (Cornell University)|Oct 1, 2014
Particle Detector Development and PerformancePhysics and Astronomy11 references4 citations
TL;DR

This paper presents an InGrid-based low-energy X-ray detector combining a Timepix pixel chip with an integrated Micromegas gas amplification stage, achieving a detection threshold as low as 277 eV and background rates below 5×10⁻⁵ /keV/cm²/s with lead shielding. The detector enables topological background suppression via high-spatial-resolution event shape analysis, demonstrating successful operation at the CERN Axion Solar Telescope (CAST).

ABSTRACT

An X-ray detector based on the combination of an integrated Micromegas stage with a pixel chip has been built in order to be installed at the CERN Axion Solar Telescope. Due to its high granularity and spatial resolution this detector allows for a topological background suppression along with a detection threshold below $1\, ext{keV}$. Tests at the CAST Detector Lab show the detector's ability to detect X-ray photons down to an energy as low as $277\, ext{eV}$. The first background data taken after the installation at the CAST experiment underline the detector's performance with an average background rate of $5 imes10^{-5}\,/ ext{keV}/ ext{cm}^2/ ext{s}$ between 2 and $10\, ext{keV}$ when using a lead shielding.

Motivation & Objective

  • To develop a low-energy X-ray detector with sub-keV threshold for axion detection at the CERN Axion Solar Telescope (CAST).
  • To reduce background rates through topological suppression using high-granularity pixel readout and event shape analysis.
  • To integrate a Micromegas gas multiplication stage directly onto a Timepix ASIC for precise alignment and improved performance.
  • To demonstrate the feasibility of a compact, robust, and highly sensitive X-ray detector in a real experimental environment.

Proposed method

  • The detector uses a Timepix ASIC with 256×256 pixels (55 µm pitch) for charge and time measurement, enabling high spatial resolution.
  • An integrated Micromegas (InGrid) stage is fabricated directly on the ASIC using photolithographic processes, with a 4 µm silicon nitride resistive layer for electrical protection.
  • The detector features a 3 cm drift gap at 500 V/cm, with a 2 µm aluminized Mylar entrance window to allow low-energy X-rays to enter while maintaining vacuum integrity.
  • A field-shaping electrode is placed above the chip to minimize electric field distortions at the chip's edges and wire bonds.
  • The detector is filled with an Ar/2.3% isobutane gas mixture to enable electron multiplication and signal readout via ionization events.
  • Background suppression is achieved using a likelihood algorithm that compares event shapes to reference distributions, enabling identification of real X-ray events based on spatial topology.

Experimental results

Research questions

  • RQ1Can an InGrid-based X-ray detector achieve a detection threshold below 1 keV, enabling the observation of low-energy X-ray lines such as the Carbon Kα at 277 eV?
  • RQ2How effective is topological background suppression using high-spatial-resolution event shape analysis in a real experimental environment with complex backgrounds?
  • RQ3What background rates can be achieved in a low-energy X-ray detection setup at CAST with proper shielding and signal discrimination?
  • RQ4Can the integration of a Micromegas stage directly on a pixel chip improve alignment precision and reduce noise compared to conventional pad-based readouts?
  • RQ5What are the operational stability and performance limits of such a detector in a long-term, vacuum-compatible setup?

Key findings

  • The detector successfully detected X-ray photons down to 277 eV, corresponding to the Carbon Kα line, confirming a sub-keV energy threshold.
  • The average background rate in the 2–10 keV range was measured at 5×10⁻⁵ /keV/cm²/s when using lead shielding, demonstrating excellent background suppression.
  • The background spectrum showed peaks attributable to copper cathode fluorescence and misidentified cosmic rays, which were effectively suppressed by the event shape analysis.
  • The likelihood-based background discrimination algorithm maintained a signal efficiency of approximately 90% across all energy ranges by optimizing cut values per energy bin.
  • The detector operated stably for two months at CAST without any detector-related failures, validating its robustness in a real experimental environment.
  • Future upgrades are planned to record the signal induced on the InGrid mesh, enabling additional event properties for further background suppression.