[Paper Review] Effects of Syringe Pump Fluctuations On Cell-Free Layer in Hydrodynamic Separation Microfluidic Devices
This study investigates how syringe pump-induced pressure fluctuations disrupt cell-free layer (CFL) stability in hydrodynamic microfluidic separation devices. Using a novel coded compressive rotating mirror (CCRM) high-speed camera, the authors demonstrate that CFL width fluctuations directly mirror pump-induced pressure oscillations in frequency and amplitude, degrading separation efficiency—especially at flow rates of 10–30 mL/h, where milk fat particles are diverted into particle-poor outlets despite nominal design.
Syringe pumps are widely used biomedical equipment which offer low-cost solutions to drive and control flow through microfluidic chips. However, they have been shown to transmit mechanical oscillations resulting from their stepper motors, into the flow, perturbing device performance. In this work, unlike previous studies at lower flow rates, we have uncovered that the relative pressure fluctuation plateau from 5mL/h onwards to approximately 2% of the average pressure. Furthermore, we find that absolute pressure fluctuations increase as a non-linear monotonic function of kinematic viscosity at flow rates in the 5-25 mL/h range, while the relative pressure fluctuations peak at 1.25 cSt. Using a novel low-cost coded compressive rotating mirror (CCRM) camera, we investigated the effect of fluctuations in a hydrodynamic microfluidic separation device based on a cell-free layer concept. Using this high-speed imaging set-up, we quantified the cell-free zone width fluctuations at bifurcations. We demonstrated that these fluctuations have the same frequency and amplitude than the syringe pump induced pressure oscillations. Finally, to illustrate that pressure fluctuations degrade the separation efficiency in such devices, we demonstrate using milk samples, instances of particles diverted to undesired outlets. This work provides an insight into the effect of syringe pump fluctuations on microfluidic separation, which will inform the design of microfluidic systems and improve their resilience to pulsating or fluctuating flow conditions.
Motivation & Objective
- To investigate the impact of syringe pump-induced pressure fluctuations on cell-free layer (CFL) stability in hydrodynamic microfluidic separation devices.
- To quantify the relationship between syringe pump mechanical oscillations and CFL width fluctuations at moderate to high flow rates (5–25 mL/h).
- To evaluate the effect of these fluctuations on separation efficiency using milk as a model two-phase sample.
- To demonstrate that pressure fluctuations degrade particle path control and lead to cross-contamination of outlets.
- To provide design guidelines for robust microfluidic systems operating with conventional syringe pumps.
Proposed method
- Employed a novel, low-cost coded compressive rotating mirror (CCRM) camera for high-speed, high-sensitivity imaging of transient CFL dynamics.
- Measured absolute and relative pressure fluctuations across flow rates from 5 to 25 mL/h using a syringe pump with a stepper motor.
- Conducted experiments with milk (50% fat) at 10, 20, and 30 mL/h to assess separation performance under fluctuating flow.
- Used absorbance spectroscopy at 220 nm and 968 nm to quantify fat content in skimmed milk and fat outlet fractions.
- Applied dynamic light scattering (DLS) to analyze particle size distribution in collected fractions.
- Performed computational simulations of particle trajectories at different depths (Z=6 μm and Z=10 μm) relative to the constriction upper boundary.
Experimental results
Research questions
- RQ1How do syringe pump-induced pressure fluctuations affect the width and stability of the cell-free layer (CFL) in microfluidic bifurcation devices?
- RQ2What is the relationship between syringe pump mechanical oscillations and CFL width fluctuations in terms of frequency and amplitude?
- RQ3How do kinematic viscosity and flow rate influence the magnitude of pressure fluctuations in the 5–25 mL/h range?
- RQ4To what extent do pressure fluctuations degrade separation efficiency in hydrodynamic microfluidic devices, as demonstrated with milk samples?
- RQ5Can the observed fluctuations in CFL width be directly linked to the mechanical oscillations of the syringe pump?
Key findings
- Relative pressure fluctuations plateau at approximately 2% of the average pressure for flow rates from 5 mL/h onwards.
- Absolute pressure fluctuations increase non-linearly with kinematic viscosity, peaking at 1.25 cSt in the 5–25 mL/h range.
- CFL width fluctuations directly mirror syringe pump-induced pressure oscillations in both frequency and amplitude, confirmed via CCRM imaging.
- At 10 mL/h, absorbance at 220 nm in the skimmed milk outlet was significantly higher (p=0.0056) than at 20 and 30 mL/h, indicating increased fat contamination due to fluctuations.
- DLS data showed a higher proportion of larger milk fat globules (1–7 µm) in the 10 mL/h skimmed milk outlet, corroborating spectrophotometric results.
- Simulations confirmed that particles located 2.5–3.5 µm from the constriction upper boundary at Z=10–6 μm could be diverted into the skimmed milk outlet under fluctuating conditions.
Read this paper