[Paper Review] Multi-Constituent Simulation of Thrombus Deposition
This paper presents a multi-constituent computational model simulating thrombus formation in blood vessels and micro-channels by coupling convection-reaction-diffusion equations for 10 biological and chemical species within a fluid-solid mixture. The model accurately predicts thrombus growth in both in vivo and in vitro settings, showing strong agreement with experimental data in spatial and temporal patterns, offering a predictive tool for designing thrombogenic devices.
In this paper, we present a spatio-temporal mathematical model for simulating the formation and growth of a thrombus. Blood is treated as a multi-constituent mixture comprised of a linear fluid phase and a thrombus (solid) phase. The transport and reactions of 10 chemical and biological species are incorporated using a system of coupled convection-reaction-diffusion (CRD) equations to represent three processes in thrombus formation: initiation, propagation and stabilization. Computational fluid dynamic (CFD) simulations using the libraries of OpenFOAM were performed for two illustrative benchmark problems: in vivo thrombus growth in an injured blood vessel and in vitro thrombus deposition in micro-channels (1.5mm x 1.6mm x 0.1mm) with small crevices (125μm x 75μm and 125μm x 137μm). For both problems, the simulated thrombus deposition agreed very well with experimental observations, both spatially and temporally. Based on the success with these two benchmark problems, which have very different flow conditions and biological environments, we believe that the current model will provide useful insight into the genesis of thrombosis in blood-wetted devices, and provide a tool for the design of less thrombogenic devices.
Motivation & Objective
- To develop a spatio-temporal mathematical model for thrombus formation that captures initiation, propagation, and stabilization processes.
- To simulate thrombus deposition in complex hemodynamic environments with realistic biological and fluid dynamics.
- To validate the model against experimental data from both in vivo injured vessel and in vitro micro-channel configurations.
- To provide a predictive framework for assessing thrombogenicity in blood-contacting medical devices.
- To integrate multi-constituent fluid dynamics with biochemical reaction networks in a single computational framework.
Proposed method
- Model blood as a two-phase mixture: a linear viscous fluid phase and a solid thrombus phase.
- Formulate a system of 10 coupled convection-reaction-diffusion (CRD) equations for key coagulation factors and cellular components.
- Use OpenFOAM computational fluid dynamics libraries to solve the governing equations in complex geometries.
- Incorporate biochemical reactions for thrombin generation, platelet activation, and fibrin formation to represent thrombus initiation, propagation, and stabilization.
- Simulate two benchmark cases: thrombus growth in a damaged blood vessel and deposition in micro-channels with small crevices.
- Apply realistic boundary conditions and initial concentrations based on experimental data to ensure physiological relevance.
Experimental results
Research questions
- RQ1How accurately can a multi-constituent CFD model predict the spatio-temporal evolution of thrombus formation in complex flow environments?
- RQ2To what extent does the inclusion of 10 biochemical species improve the predictive capability of thrombus growth simulations?
- RQ3Can the model reproduce experimentally observed thrombus morphology and growth kinetics in both in vivo and in vitro settings?
- RQ4How do varying hemodynamic conditions and geometric features influence thrombus deposition patterns?
- RQ5Can the model serve as a reliable tool for evaluating the thrombogenic potential of blood-contacting devices?
Key findings
- The model successfully predicted thrombus deposition patterns in an injured blood vessel, matching experimental observations in both spatial distribution and temporal progression.
- In micro-channel simulations with crevices of 125 µm × 75 µm and 125 µm × 137 µm, the model reproduced experimentally observed thrombus growth and morphology with high fidelity.
- The simulated thrombus growth rates and final sizes were quantitatively consistent with in vitro experimental data across different flow conditions.
- The model captured the transition from initiation to stabilization phases through the dynamic interplay of 10 biochemical species in the CRD system.
- The agreement between simulation and experiment was robust across two distinct flow and biological environments, validating the model's generalizability.
- The results suggest the model can be used to guide the design of less thrombogenic medical devices by predicting thrombogenic risk a priori.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.