[Paper Review] Phase-Aberration Correction in Shear-wave Elastography Imaging Using Local Speed-of-Sound Adaptive Beamforming
This study proposes local speed-of-sound (SoS) adaptive beamforming to correct phase aberrations in shear-wave elastography imaging (SWEI), which otherwise introduce significant bias in shear-wave speed (SWS) estimation. Using simulations and ex-vivo porcine muscle phantoms, the authors demonstrate that beamforming with a known local SoS map reduces SWS estimation disparity by over 4.3 times compared to constant-SoS beamforming, proving that accurate SoS distribution is critical for reliable elasticity quantification.
Shear-wave Elastography Imaging (SWEI) is a noninvasive imaging modality that provides tissue elasticity information by measuring the travelling speed of an induced shear-wave. It is commercially available on clinical ultrasound scanners and popularly used in the diagnosis and staging of liver disease and breast cancer. In conventional SWEI methods, a sequence of acoustic radiation force (ARF) pushes are used for inducing a shear-wave, which is tracked using high frame-rate multi-angle plane wave imaging (MA-PWI) to estimate the shear-wave speed (SWS). Conventionally, these plane waves are beamformed using a constant speed-of-sound (SoS), assuming an a-priori known and homogeneous tissue medium. However, soft tissues are inhomogeneous, with intrinsic SoS variations. In this work, study the SoS effects and inhomogeneities on SWS estimation, using simulation and phantoms experiments with porcine muscle as an abbarator, and show how these aberrations can be corrected using local speed-of-sound adaptive beamforming. For shear-wave tracking, we compare standard beamform with spatially constant SoS values to software beamforming with locally varying SoS maps. We show that, given SoS aberrations, traditional beamforming using a constant SoS, regardless of the utilized SoS value, introduces a substantial bias in resulting SWS estimations. Average SWS estimation disparity for the same material was observed over 4.3 times worse when a constant SoS value is used compared to that when a known SoS map is used for beamforming. Such biases are shown to be corrected by using a local SoS map in beamforming, indicating the importance of and a need for local SoS reconstruction techniques.
Motivation & Objective
- To investigate the impact of speed-of-sound (SoS) inhomogeneities on shear-wave speed (SWS) estimation in shear-wave elastography imaging (SWEI).
- To evaluate the bias introduced in SWS estimation when using a constant SoS value for beamforming in multi-angle plane wave imaging (MA-PWI).
- To demonstrate that local SoS-adaptive beamforming can correct phase aberrations and reduce SWS estimation errors.
- To establish the necessity of accurate local SoS reconstruction for reliable tissue elasticity quantification in clinical SWEI.
Proposed method
- Simulated shear-wave propagation through layered media with varying SoS (water and porcine muscle) to model inhomogeneous soft tissues.
- Ex-vivo experiments using porcine muscle phantoms to validate simulation results under controlled conditions.
- Beamforming of multi-angle plane wave (MA-PWI) data using both constant SoS and locally varying SoS maps derived from prior knowledge.
- SWS estimation via cross-correlation of displacement time profiles (DTPs) between consecutive frames to track shear-wave front propagation.
- Comparison of SWS estimates using constant-SoS beamforming versus SoS-adaptive beamforming with spatially varying SoS maps.
- Use of direct linear path beamforming without refraction modeling to isolate SoS-related beamforming errors.
Experimental results
Research questions
- RQ1How do SoS inhomogeneities in soft tissues affect shear-wave speed (SWS) estimation in SWEI when using constant-SoS beamforming?
- RQ2Can local speed-of-sound adaptive beamforming reduce or eliminate SWS estimation bias caused by phase aberrations?
- RQ3Is there a single constant SoS value that can mitigate SWS estimation errors across heterogeneous media?
- RQ4To what extent does beamforming with a known local SoS map improve SWS estimation accuracy compared to constant-SoS beamforming?
- RQ5What is the magnitude of SWS estimation disparity when using incorrect or constant SoS values in beamforming for SWEI?
Key findings
- Using a constant SoS value for beamforming results in an average SWS estimation disparity over 4.3 times worse than using a known local SoS map.
- No single constant SoS value could eliminate the SWS estimation bias across different tissue layers, even when the true SoS was known.
- SWS estimation errors were primarily due to beamforming inaccuracies rather than shear-wave generation, confirming that beamforming is the main source of bias.
- The use of a local SoS map in beamforming eliminated systematic SWS estimation bias, even when refraction effects were not modeled.
- For the same tissue, SWS-based elastic modulus estimates varied by up to 3.89 kPa (from 4.67 kPa to 8.94 kPa) depending on the beamforming SoS assumption, highlighting clinical risk of misdiagnosis.
- The results indicate that accurate local SoS reconstruction is essential for standardizing SWEI measurements and ensuring reliable disease staging.