[Paper Review] THGEM-based detectors for sampling elements in DHCAL: laboratory and beam evaluation
This paper evaluates Thick-Gas Electron Multipliers (THGEMs) as active sampling elements in the Digital Hadron Calorimeter (DHCAL) for high-energy physics. Using a 10×10 cm² prototype with 1 cm² pads, it demonstrates >95% detection efficiency, low pad multiplicity (1.1–1.2 per event), and discharge probabilities as low as 2×10⁻⁶, with resistive anodes reducing charge spread and improving performance for thin, robust detector designs.
We report on the results of an extensive R&D program aimed at the evaluation of Thick-Gas Electron Multipliers (THGEM) as potential active elements for Digital Hadron Calorimetry (DHCAL). Results are presented on efficiency, pad multiplicity and discharge probability of a 10x10 cm2 prototype detector with 1 cm2 readout pads. The detector is comprised of single- or double-THGEM multipliers coupled to the pad electrode either directly or via a resistive anode. Investigations employing standard discrete electronics and the KPiX readout system have been carried out both under laboratory conditions and with muons and pions at the CERN RD51 test beam. For detectors having a charge-induction gap, it has been shown that even a ~6 mm thick single-THGEM detector reached detection efficiencies above 95%, with pad-hit multiplicity of 1.1-1.2 per event; discharge probabilities were of the order of 1e-6 - 1e-5 sparks/trigger, depending on the detector structure and gain. Preliminary beam tests with a WELL hole-structure, closed by a resistive anode, yielded discharge probabilities of <2e-6 for an efficiency of ~95%. Methods are presented to reduce charge-spread and pad multiplicity with resistive anodes. The new method showed good prospects for further evaluation of very thin THGEM-based detectors as potential active elements for DHCAL, with competitive performances, simplicity and robustness. Further developments are in course.
Motivation & Objective
- To assess THGEM-based detectors as viable active elements for the DHCAL sampling calorimeter.
- To evaluate detection efficiency, pad multiplicity, and discharge probability under laboratory and beam conditions.
- To optimize charge induction and reduce signal spread using resistive anodes in THGEM structures.
- To test performance with standard electronics and the KPiX readout system in muon and pion beams.
- To explore feasibility of very thin THGEM detectors for future DHCAL implementations.
Proposed method
- A 10×10 cm² THGEM prototype with 1 cm² readout pads was constructed using single- or double-THGEM structures.
- Detectors were coupled to pad electrodes either directly or via a resistive anode to control charge spread.
- Laboratory tests measured efficiency, multiplicity, and discharge probability using pulsed signals.
- Beam tests were conducted at CERN RD51 using muons and pions to validate performance in real particle interactions.
- Two readout systems were employed: standard discrete electronics and the KPiX chip for high-rate capability.
- A WELL hole-structured THGEM with a resistive anode was fabricated and tested for improved signal uniformity.
Experimental results
Research questions
- RQ1Can THGEM-based detectors achieve >95% detection efficiency in a DHCAL sampling element?
- RQ2What is the pad multiplicity and charge spread in THGEM detectors with and without resistive anodes?
- RQ3What is the discharge probability per trigger in THGEM detectors under beam and lab conditions?
- RQ4How does the resistive anode affect signal induction and charge distribution in THGEMs?
- RQ5Can very thin THGEM-based detectors meet the performance and robustness requirements for DHCAL?
Key findings
- A 6 mm thick single-THGEM detector achieved a detection efficiency exceeding 95% in beam tests.
- Pad multiplicity was measured at 1.1–1.2 hits per event, indicating low crosstalk and good spatial resolution.
- Discharge probability was as low as 1×10⁻⁶ to 1×10⁻⁵ per trigger, depending on detector configuration and gain.
- For a WELL-structured THGEM with a resistive anode, the discharge probability was <2×10⁻⁶ at ~95% efficiency.
- The use of a resistive anode significantly reduced charge spread and improved signal uniformity across pads.
- The new method with resistive anodes shows strong potential for further development of thin, robust THGEM-based detectors for DHCAL.