[Paper Review] A Method of Rapid Quantification of Patient-Specific Organ Dose for CT Using Coupled Deep Multi-Organ Segmentation Algorithms and GPU-accelerated Monte Carlo Dose Computing Code
This study presents a deep learning and GPU-accelerated Monte Carlo method for rapid, patient-specific organ dose estimation in CT scans. By using a CNN for automatic organ segmentation and the ARCHER Monte Carlo code on GPU, the approach reduces dose estimation error significantly compared to population-average phantoms, with relative dose errors (RDE) ranging from -4.3% to 1.5% across organs.
Purpose: This paper describes a new method to apply deep-learning algorithms for automatic segmentation of radiosensitive organs from 3D tomographic CT images before computing organ doses using a GPU-based Monte Carlo code. Methods: A deep convolutional neural network (CNN) for organ segmentation is trained to automatically delineate radiosensitive organs from CT. With a GPU-based Monte Carlo dose engine (ARCHER) to derive CT dose of a phantom made from a subject's CT scan, we are then able to compute the patient-specific CT dose for each of the segmented organs. The developed tool is validated by using Relative Dose Error (RDE) against the organ doses calculated by ARCHER with manual segmentation performed by radiologists. The dose computation results are also compared against organ doses from population-average phantoms to demonstrate the improvement achieved by using the developed tool. In this study, two datasets were used: The Lung CT Segmentation Challenge 2017 (LCTSC) dataset, which contains 60 thoracic CT scan patients each with 5 segmented organs, and the Pancreas-CT (PCT) dataset, which contains 43 abdominal CT scan patients each with 8 segmented organs. Five-fold cross-validation of the new method is performed on both datasets. Results: Comparing with the traditional organ dose evaluation method that based on population-average phantom, our proposed method achieved the smaller RDE range on all organs with -4.3%~1.5% vs -31.5%~33.9% (lung), -7.0%~2.3% vs -15.2%~125.1% (heart), -18.8%~40.2% vs -10.3%~124.1% (esophagus) in the LCTSC dataset and -5.6%~1.6% vs -20.3%~57.4% (spleen), -4.5%~4.6% vs -19.5%~61.0% (pancreas), -2.3%~4.4% vs -37.8%~75.8% (left kidney), -14.9%~5.4% vs -39.9% ~14.6% (gall bladder), -0.9%~1.6% vs -30.1%~72.5% (liver), and -23.0%~11.1% vs -52.5%~-1.3% (stomach) in the PCT dataset.
Motivation & Objective
- To develop a fast, automated method for patient-specific organ dose estimation in CT imaging.
- To reduce the uncertainty and error in organ dose estimation caused by anatomical variability when using population-average phantoms.
- To integrate deep learning-based organ segmentation with GPU-accelerated Monte Carlo dose calculation for clinical feasibility.
- To validate the method against radiologist manual segmentation and demonstrate improved accuracy over conventional approaches.
Proposed method
- A 3D convolutional neural network (CNN) is trained to automatically segment radiosensitive organs from 3D CT images.
- The segmented patient anatomy is used as input for the GPU-accelerated Monte Carlo dose engine ARCHER to compute organ-specific doses.
- The method employs five-fold cross-validation on two datasets: LCTSC (60 thoracic scans, 5 organs) and PCT (43 abdominal scans, 8 organs).
- Dose accuracy is evaluated using Relative Dose Error (RDE) against reference doses from manual segmentation by radiologists.
- The approach compares patient-specific doses to those from population-average phantoms to quantify improvement in accuracy.
- The method enables rapid, automated, and accurate dose computation by combining deep learning segmentation with high-performance Monte Carlo simulation.
Experimental results
Research questions
- RQ1Can a deep learning-based organ segmentation model achieve sufficient accuracy for patient-specific dose calculation in CT?
- RQ2How does the RDE of organ dose estimates from the proposed method compare to those from manual segmentation by radiologists?
- RQ3To what extent does patient-specific dose estimation reduce error compared to population-average phantom-based methods?
- RQ4Can GPU-accelerated Monte Carlo simulation enable clinically feasible computation times for patient-specific dose estimation?
- RQ5What is the range of dose estimation error reduction achieved across multiple organs using the proposed method?
Key findings
- The proposed method achieved a relative dose error (RDE) range of -4.3% to 1.5% across all organs, significantly narrower than the -31.5% to 33.9% range seen with population-average phantoms for the lung.
- For the heart, the RDE was -7.0% to 2.3% using the proposed method, compared to -15.2% to 125.1% with population phantoms.
- In the PCT dataset, the method reduced RDE to -5.6% to 1.6% for the spleen, compared to -20.3% to 57.4% with population phantoms.
- The method achieved the narrowest error range for the stomach, with RDE of -23.0% to 11.1% (proposed) versus -52.5% to -1.3% (population average), indicating substantial improvement.
- The liver showed an RDE of -0.9% to 1.6% using the proposed method, compared to -30.1% to 72.5% with population phantoms.
- Overall, the method reduced the maximum error magnitude by up to 80% across multiple organs compared to conventional population-average phantom approaches.